SYSTEM_FAULT_COM_ERROR
SYSTEM_FAULT_SYSTEM_ERROR
SYSTEM_FAULT_CHANNEL_ERROR
SYSTEM_FAULT_NOVALUM_HAL_ERROR
SYSTEM_FAULT_SYSCOM_HAL_ERROR
SYSTEM_FAULT_INTERLOCK_CVR_REED_OPEN
SYSTEM_FAULT_OVER_FREQUENCY
SYSTEM_FAULT_I2C_TIMEOUT
15000-MNL-01 Rev A
range. Please contact a technical
representative if this error persists.
Communications Error.
Check USB cable connection to the host
PC. Unplug the unit and wait 10 seconds
before reapplying power to the system. If
the error persists, please contact a
technical representative.
System Error.
Unplug the unit and wait 10 seconds
before reapplying power to the system. If
the error persists, please contact a
technical representative.
Channel Error.
Unplug the unit and wait 10 seconds
before reapplying power to the system. If
the error persists, please contact a
technical representative.
Novalum HAL Error.
Unplug the unit and wait 10 seconds
before reapplying power to the system. If
the error persists, please contact a
technical representative.
Interlock HAL Error.
Unplug the unit and wait 10 seconds
before reapplying power to the system. If
the error persists, please contact a
technical representative.
The Fiber Access Cover is
Verify that the fiber access cover is
open.
completely closed, and nothing is in-
between the cover and the LDI. See
Figure 12.
Over-Frequency.
Decrease the frequency at which the
laser is being pulsed.
I2C Timeout.
Unplug the unit and wait 10 seconds
before reapplying power to the system. If