89 North LDI-6 Operating Manual page 29

Table of Contents

Advertisement

SYSTEM_FAULT_COVER_REMOVED
SYSTEM_FAULT_INTERLOCK_USER_OPEN
SYSTEM_FAULT_INTERLOCK_FIBER1_OPEN
SYSTEM_FAULT_INTERLOCK_FIBER2_OPEN
SYSTEM_FAULT_OVER_CURRENT
SYSTEM_FAULT_NECSEL_OVER_TEMP
SYSTEM_FAULT_NECSEL_UNDER_TEMP
SYSTEM_FAULT_NOVALUM_OVER_TEMP
SYSTEM_FAULT_NOVALUM_UNDER_TEMP
15000-MNL-01 Rev A
Interlock - Cover
Please contact a technical representative
Removed.
if this error persists.
Interlock - Remote Open.
Verify that the eyepiece interlock on the
back of the LDI is properly installed and in
the proper state.
Interlock - Novalum Fiber
Verify that the interlock switch in the
RGGBBV fiber routing slot on the front of
Open.
the LDI is completely depressed and the
"Laser Ready" light is blue. See Figure 11.
Interlock - 555 Fiber Open. Verify that the interlock switch in the 555
fiber routing slot on the front of the LDI is
completely depressed and the "Laser
Ready" light is blue. See Figure 11.
Over current.
Please contact a technical representative
if this error persists.
Necsel Over Temp.
Close all shutters and let the system cool
down. If the error persists, verify that the
unit is operating within the specified
ambient temperature range and nothing
is blocking the air inlet or exhaust.
Necsel Under Temp.
Verify that the unit is operating within
the specified ambient temperature
range. Please contact a technical
representative if this error persists.
Close all shutters and let the system cool
Novalum Over Temp.
down. If the error persists, verify that the
unit is operating within the specified
ambient temperature range and nothing
is blocking the air inlet or exhaust.
Novalum Under Temp.
Verify that the unit is operating within
the specified ambient temperature
29