Table of Contents

Advertisement

Industrial Metal Detectors
THS/21E & THS/21
Document code
FI002K0021v1100UK
Read this Manual carefully before installing, operating or carrying out maintenance on the device. Keep the Manual in a safe
place for future reference, and in perfect condition. This Manual must accompany the device described therein in the
case of change of ownership, and until the device is broken up.
FI002K0021v1100UK – THS/21E and THS/21 Programming Manual
Programming Manual
Date
20/12/2011
Software version
THSV5401 – ALMV5400
1

Advertisement

Table of Contents
loading

Summary of Contents for Ceia THS/21E

  • Page 1 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual Industrial Metal Detectors THS/21E & THS/21 Programming Manual Document code Date Software version FI002K0021v1100UK 20/12/2011 THSV5401 – ALMV5400 Read this Manual carefully before installing, operating or carrying out maintenance on the device. Keep the Manual in a safe place for future reference, and in perfect condition.
  • Page 2 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual SYMBOLS The equipment is marked with this symbol wherever the user should refer to this Manual in order to avoid possible damage. The same symbol appears in the Manual at points where warnings or particularly important instructions, essential for safe, correct operation of the device, are given.
  • Page 3: Table Of Contents

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual CONTENTS     SAFETY INSTRUCTIONS – WARNINGS ..5     PROGRAMMING PARAMETERS ....15     General warnings ........5     Administrator .......... 15     Correct use of the device ......5  ...
  • Page 4 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual     3.23 Access via Web Server ......78     APPENDIX ............. 89     3.23.1 Network configuration ......78     Parameters list – Alphabetical order ..89     3.23.2 Functions available ........
  • Page 5: Safety Instructions - Warnings

    1.2 Correct use of the device The Metal Detectors from the THS/21E and THS/21 series are electronic devices for the detection of metal masses transiting inside the detection antenna. The final user is responsible for selecting the appropriate sensitivity for their application.
  • Page 6: Programming

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING 2 PROGRAMMING The list of parameters and menus, accessible by each users, together with the system behaviors, can change in respect of this Manual description, depending on the settings modifications. Contact the System Administrator for any information.
  • Page 7: Remote Communication

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING 2.1.1.2 Remote Communication For the Remote Communication, two access type are provided: Only the password (predefined password is 000006). This access type is selected by default. User and Password exactly as a local access (see table above).
  • Page 8: General Points On Programming

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING 2.4 General points on programming 2.4.1 Keypad functions The control of the Metal Detector and the setting of the device parameters are performed through the control panel keyboard, as shown below: On THS/xx21 and THS/xx21E...
  • Page 9: Entering The Username

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING 2.4.2 Entering the username In the following example, the way to digit the username “STEVEN” is shown. Display Keys pressed Meaning User Press P to enter Programming. The cursor • • #_____ blinks in the position of the first character Press four times the key “7”...
  • Page 10: Entering The Password

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING 2.4.3 Entering the password Display Keys pressed Meaning After the user name, the password is • • Password required. As an example we will digit #_____ “000004”. Push “0” to get for the letter “0” as the first •...
  • Page 11: Exit From Programming

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING 2.5 Exit from programming 2.5.1 Exit from programming Press the P key to exit from the current menu (or submenu), until the main screenshot is shown on the display. 2.5.2 Time out The programming session will be terminated automatically if no commands are entered for a period of 5 minutes.
  • Page 12: Quick Access Menu Programming Procedure

    PROGRAMMING 2.7 Quick Access menu programming procedure This function is not available on THS/21E models. The system Administrator can program up to 10 shortcuts to menus or single parameters, in order to quickly activate them through the Quick Access and the correspondent numeric key (0-9).
  • Page 13: Programming The Metal Detector According To The Kind Of Product

    Min.trans.number 1 not sure about the correct detection mode of the product. On THS/21E-3F set Autolearn > Multi-band to ON to perform an autolearn for all bands. Set it to OFF to perform it only on the current band. Set it to ON if not sure about the correct band of the product.
  • Page 14 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING Display THS/xx21E Display THS/xx21 Step PASS PRODUCT Pass the non-contaminated product through the Metal PASS PRODUCT Prod. PROD1 Detector each time is requested. The message PASS PRODUCT appears on the display, and the buzzer (if enabled) is activated.
  • Page 15: Programming Parameters

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3 PROGRAMMING PARAMETERS Administrator After user logging, the list of available Local Programming Menus is Reset Product selection shown. The list changes according to the permission levels of the Products Autolearn user logged.
  • Page 16: Remote Passwords

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Erase User 6 alphanumeric chars. Global Description Models All models Delete an user THS/xx21E THS/xx21 Scroll the users list through the arrow keys. Press E to...
  • Page 17: Reset

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.2 Reset Remote Possible settings Standard setting Parameter Reset Global Description Models Alarm and fault Reset This command restore the system from alarms and faults that require a Reset operation. All models RS command does not reset the Emergency condition: in this case, the Emergency button need to be released and the R button, if present, needs to be pressed (see Installation Manual).
  • Page 18: Products

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.4 Products Product List of commands available under the Products menu Rename product New product Erase product Prod. Barcode Save prod.to USB Load prd from USB Remote Possible settings Standard setting...
  • Page 19: New Product

    Creation of a new product Create a new product. The maximum number of products is 500 for Product SUGAR01 All models Rename product the THS/21 series and 250 for the THS/21E series. [\\\ New product \\\] BISC#_______ {|||||||||||||||||||} THS/xx21E THS/xx21...
  • Page 20: Save Prod.to Usb

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.4.1 Save prod.to USB Select product(s) List of commands available under the Save prod.to USB submenu. Save all products This menu is available only on models with IXC card installed. Remote...
  • Page 21: Save All Products

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Save all products Global Description Models Save all Products to USB memory All models This command is available only if IXC=ON THS/xx21E THS/xx21 If an USB key is not already inserted on the Metal...
  • Page 22: Load Prd. From Usb

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.4.2 Load prd. from USB This menu is available only on models with IXC card installed. THS/xx21E THS/xx21 [\ Select directory \] Enter in this Menu and select the Metal Detector to copy...
  • Page 23: Load All

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Load all Global Description Models All models Load all Products settings from USB memory This command is available only if IXC=ON THS/xx21E THS/xx21 If an USB key is not already inserted on the Metal...
  • Page 24: Autolearn

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.5 Autolearn Autolearn List of commands available under the Autolearn menu. Aut.det.mode sel Multi-band Min.transits numb Remote Possible settings Standard setting Parameter Autolearn Global ON / OFF Description Models Product autolearn This command activates the Autolearn procedure, in which the Metal Detector set automatically the best Product settings for the “product effect”...
  • Page 25: Autolearn-Advanc

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.6 Autolearn-advanc Metal optimiz. List of commands available under the Autolearn-advanc menu Autolearn sens. Sens. margin Max learn sens Vibr.immunity Ej.during autlr Remote Possible settings Standard setting Parameter Metal optimiz. AVRG / BFE...
  • Page 26: Detection

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Vibr.immunity Global 0 – 250 Description Models Vibration immunity This parameter has effect only after the Autolearn execution. All models Setting VI between 1 and 250, the Autolearn procedure optimizes the Product settings both for the except best detection sensitivity and against possible mechanical vibrations.
  • Page 27: Tx Program

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Related to the TX program 0 - 17999 current product Description Models Transmission Program Only on the best minimization for the product effect and it is automatically...
  • Page 28: Ejection

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.8 Ejection Ejection mode List of commands available under the Ejection menu Ej. distance Ejection time Pack length Ej.st.sync.area Ej.en.sync.area Remote Possible settings Standard setting Parameter Related to the B / F / S / SB / R...
  • Page 29 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Ejection Description Models mode Usable only with flat Reversed belts. No photocell required belts, capable of Used for bulk product reversed running All models In case of Alarm, the conveyor belt direction is reversed for a distance ED.
  • Page 30: Ej. Distance

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Related to the Ej. distance 0 – 6000 mm current product Description Models Distance of ejection system from probe or photocell This parameter indicates the position of the ejection system and it should be set according to the Ejection mode (EM parameter) as shown below.
  • Page 31: Ej.st.sync.area

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Related to the Ej.st.sync.area 0 – 250 mm current product Description Models Ejection start synchronization area Set ES (typically equal to PLEN) in order to synchronize the ejection start with the beginning of the All models pack in transit.
  • Page 32: Counters

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.9 Counters Prd alarms List of commands available under the Counters menu. Prd packs Tot.alarms Tot. packs Work time Remote Possible settings Standard setting Parameter Related to the Prd alarms 0 – 999999999...
  • Page 33: Configuration

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Work time Global Description Models Working time All models This parameter returns the number of operative hours of the Metal Detector. It is not possible to reset this counter.
  • Page 34: Belt Length

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Belt length BLEN Global 0 – 20000 mm Description Models Length of the conveyor belt All models Set BLEN value to the belt total length (development). Set BLEN=0 in systems without belt.
  • Page 35: Keyb Alarm Reset

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Keyb alarm reset Global ON / OFF Description Models Enabling of the keyboard for direct alarm reset This parameter set the Reset mode when manual Alarm Reset is required (i.e. after an Alarm with EM=B).
  • Page 36: Ejection

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Ejection Global ON / OFF Description Models Ejection disabling/enabling Enabling/disabling of the ejection system according to the Ejection Mode (see EM parameter). All models If disabled, shall be enabled during Metal Detector tests: a message will be shown on the display.
  • Page 37: Configuration-Adv

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.11 Configuration-adv IXC presence List of parameters available under the Configuration-adv menu. Backup conf.to USB Restor.cfg.from USB Export conf.to USB Import cfg.from USB Network config. Product tracking Inhib. time Barcode enable...
  • Page 38 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Backup conf.to USB Global Description Models All models Metal Detector Configuration Backup This command is available only if IXC=ON THS/xx21E THS/xx21 If an USB key is not already inserted on the Metal...
  • Page 39 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Export conf.to USB Global Description Models All models Export the Metal Detector Configuration for a similar Metal Detector system This command is available only if IXC=ON...
  • Page 40: Network Config

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.11.1 Network config. HN~~~~~THS2110021111@ List of commands available under the Network config. submenu. Host protocol STAT 192. 9.201. 96 255.255.255. 192. 9.201.100 This menu is available only on models with IXC card installed 192.168.
  • Page 41 Description Models Enabling code for operation with bar-code reader 6-character code supplied by CEIA and specific for each Unit. All models This code enables the Metal Detector to work with a Bar-Code reader. Settings BE with the right code enable the visualization of the Barcode Menu with all the corresponding parameters.
  • Page 42 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Related to the Speed 2 – 250 m/min. current product Description Models Product transit speed This parameter set the nominal value for the product transit speed [m/min].
  • Page 43 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Check phot.dist ECPD Global 0 – 6000 mm Description Models Distance between Metal Detector and Ejection Check photocell Set ECPD to the distance between the center of the Ejection Check Photocell and the exit side of All models the Probe.
  • Page 44 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter TP adjustment Global 0 – 35999 Description Models Correction of the received analysis signal Only on This parameter is specific for each device and automatically set during TA Autolearn.
  • Page 45 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Autotest module ATTM Global ON / OFF Description Models All models, Autotest signal Magnitude tracking except Automatic control, continuously performed, to guarantee the maximum Metal Detector operative THS/MN21 and stability in every condition.
  • Page 46 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter BIN FULL input Global NC / NO Description Models Function type of the device connected to the “BIN FULL” input NC: Normally Closed; NO: Normally Open...
  • Page 47 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter LOW PRESSURE inp Global NC / NO Description Models Function type of the device connected to the "LOW PRESSURE" input NC: Normally Closed; NO: Normally Open All models Set NC or NO according to the output type of the connected sensor, in normal working condition.
  • Page 48 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Eject.act.if stop Global ON / OFF Description Models Ejector activation when belt stops With EIS=ON the system has different behaviours, according to its configuration: - With Conveyor Control System or encoder: in case of a belt stop during an ejection, the ejector will continue to be activated until the belt restart and then it will finish the ejection.
  • Page 49 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Autom. restart Global ON / OFF Description Models Belt restart if there is “FOLLOWING CONVEYOR” input activation If FC= OFF, the parameter has no effect;...
  • Page 50: Motor Config

    All models This parameter is adjustable only with the motor stopped. Available only with Conveyor Control System On CEIA integrated systems the value is already set to ensure the correct functioning of the system itself Remote Possible settings Standard setting Parameter Nom.MotorCurr.
  • Page 51 All models This parameter is adjustable only with the motor stopped. Available only with Conveyor Control System On CEIA integrated systems the value is already set to ensure the correct functioning of the system itself Remote Possible settings Standard setting Parameter Nom.Motor.Speed...
  • Page 52 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter 2, 4, 6, 8, 10, Inv.PulseFreq. Global 12, 14, 16 kHz Description Models PWM frequency All models This parameter is adjustable only with the motor stopped.
  • Page 53: Barcode Reader

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Bounce t. slow Global 0,001 – 0,200 s 0,05 Description Models Setting of the minimum activation time for slow inputs All models Minimum activation time for a “slow” input (control keys and emergency buttons), to be considered active.
  • Page 54 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Learn code Global ON / OFF Description Models Activation of code/product association phase Command which allows a code to be associated with a product. When activated (LC=ON), the message “Pass Code”...
  • Page 55: I/O Status

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter No code ejection Global ON / OFF Description Models Ejection of packs with wrong code When activated, the packages without barcodes or with invalid barcodes are ejected. If deactivated, All models there is no action.
  • Page 56: Measures

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.13.3 Measures Speed This submenu shows a list of measures performed in real time by the Motor curr. Motor volt V11 Metal Detector. This can be useful during Technical Assistance. ALM supply V1 24.0...
  • Page 57: Diagnosis Managem

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.14 Diagnosis managem. Photocell fault List of parameters available under the Diagnosis managem. menu. Sync.max.activ Chck.max.activ Phot.fau.thres Bin full Bin full delay Bin absent Bin absent delay Bin with lock Unlock bin time Air pressure low Low press.
  • Page 58 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Bin full Global ON / OFF / CEON Description Models Bin Full fault activation Set LF = ON to enable the “Bin Full” Fault activation in case of BIN FULL Input activated for a time longer than LFD delay.
  • Page 59 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Air pressure low Global ON / OFF Description Models Low Pressure fault activation Set AP = ON to enable the “Low Pressure” Fault activation in case of LOW PRESSURE Input All models activated for a time longer than APDL delay.
  • Page 60 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Ej.check accur. ECKA Global 2 – 250 mm Description Models Accuracy of the ejection check functionality Setting ECKA: EM=FS: the area to be checked is equal to the area cleared by the ejector, reduced of ECKA value All models on both sides.
  • Page 61 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Test time out Global ON / OFF Description Models Activation of fault status if maximum time for Test expires Set FT=ON to enable the Fault ”Test Timeout” activation in case of a scheduled Test not started All models within the maximum time (TD or FTD).
  • Page 62 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Al.rate threshold Global 0 – 250 Description Models Setting of alarm rate threshold Set ART > 0 to enable the control of the Alarms number, according to ARP parameter: ARP=0: if the number of current product Alarms (CA) reaches ART value, the Fault “Too many...
  • Page 63: Md Test

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.15 MD Test FExtestx List of commands available under the MD Test menu. NFExtest SSxtestx Remote Possible settings Standard setting Parameter Related to the FE test current product Description Models Activation of the test with Ferrous metal Test Sample Select the test and press E key to start it.
  • Page 64: Print

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.17 Print The list of commands available under the Print menu changes accordingly with the presence of an IXC card. 3.17.1 With no IXC card installed Print last events List of commands available under the Print menu, when no IXC card...
  • Page 65: With Ixc Card Installed

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Erase printed ev Global Description Models Erasing of printed events This command confirm the events printed with PL, from the last EV command and consequently All models make available the corresponding memory space for next events recordings.
  • Page 66: Save Events

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.17.2.1 Save events Save from-to List of commands available under the Save events submenu. Save all events Remote Possible settings Standard setting Parameter Save from-to Global Description Models All models Data saving of events occurred in a time period...
  • Page 67 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Save all events Global Description Models All models Data saving of all events THS/xx21E THS/xx21 If an USB key is not already inserted on the Metal Detector...
  • Page 68: Q-C Setup (Quality Control)

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.18 Q-C setup (Quality Control) 1st test delay List of parameters available under the Q-C setup menu. Test period Test max delay Test mode Sample ampl.check Quick test Succ.test mandat Remote...
  • Page 69 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter OFF, X2, X3, X4, Sample ampl. check Global X6, X8, X12, X16, X24, X32 Description Models Sample Amplitude Check If SAC = “Xv” value, during each test the Metal Detector checks that the Test sample signal level will not be “v”...
  • Page 70: Test Samples

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.19 Test samples FE diameter List of parameters available under the Test samples menu. NFE diameter SS diameter Remote Possible settings Standard setting Parameter Related to the FE diameter 0.0 – 25.0 mm...
  • Page 71: Commands Accessible Only In Remote Programming

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.21 Commands accessible only in Remote Programming Remote Possible settings Standard setting Parameter Global Description Models Exit from remote programming Type this command to exit from the Remote Programming. All models If RPTO=ON, the Metal Detector exit automatically from the Remote Programming in case of 5 minutes of programming inactivity.
  • Page 72 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Related to current product Description Models List of all current product parameters and their value All models List of parameters related to the current Product, together with their actual value.
  • Page 73 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Global Description Models Display and Alarm Led status The MD replies to this command with “DL(SP)” followed by a 81-character string, (CR), (CR) and ‘#’.
  • Page 74 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Global Description Self-Check This command returns a list of all the Metal Detector status flag which are active. The Metal Detector will reply to this command with “SC xx,xx,…..,xx(CR)(CR)#” where “xx” are ASCII strings with the...
  • Page 75 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Remote Possible settings Standard setting Parameter Global Description Models Exit from programming When you send this command to the THS Metal Detector, the screen will jump directly to TOP All models screen from anywhere in the programming tree.
  • Page 76: Remote Programming

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.22 Remote programming Programming can be carried out through a remote computer fitted with an RS232 interface, a Bluetooth connection, or an Ethernet/WiFi connection, depending on their availability on the Metal Detector, using a standard communications program (e.g. Windows HyperTerminal) or the CEIA MDScope program, available on request.
  • Page 77: Entering Remote Programming

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.22.2 Entering remote programming 3.22.2.1 Connecting to a Metal Detector Via RS232 or Bluetooth Run the serial communication program (MD Scope) or another serial communication software Perform the access by typing the password...
  • Page 78: Access Via Web Server

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.23 Access via Web Server This paragraph applies only if a IXC card is fitted. 3.23.1 Network configuration To configure the THS system on a network, use the following procedure: Enter the programming phase as Administrator;...
  • Page 79: Status

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.23.2.1 Status This page replicates every message the THS shows on the display. The page is updated each 5 s. 3.23.2.2 Products This page shows the list of product saved into the THS memory, showing the selected product in...
  • Page 80: Products Counters

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.23.2.3 Products Counters This page shows, in real time, the alarm counters status. 3.23.2.4 Test Diameters This page shows the diameter values of all the test samples.
  • Page 81: Products Export

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.23.2.5 Products Export From this page it is possible to select all or some of the products stored in the THS memory, to export them. Each selected product will be saved in a ProductName.TPR file and all of them will be saved into a Exported_Products.zip file.
  • Page 82: Products Delete

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.23.2.7 Products Delete If needed, from this page it is possible to delete products. 3.23.2.8 Buffers Export Inside this page it is possible to export events data that occurred in from the Starting Date / Time to the Ending Date / Time requested.
  • Page 83: Configuration Backup/Restore

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.23.2.9 Configuration Backup/Restore From this page it is possible to backup or to restore the configuration for the connected system. Configuration Backup To perform a configuration backup, simply press on Click here to Backup current configuration.
  • Page 84: Configuration Export/Import

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.23.2.10 Configuration Export/Import From this page it is possible to export the configuration to or to import it from other Metal Detectors of the same model, installed on identical systems. It is possible to import configuration from Metal Detectors of the same model, installed on identical systems, but with different probe dimensions.
  • Page 85: Network

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.23.2.11 Network This function allows to set all network data of the THS system. When using DHCP protocol, the IP address will be assigned automatically by the network system. Otherwise, using a static addressing, the network data must be entered manually.
  • Page 86: System Upgrade (Web Server)

    This function allows performing some maintenance operations. System Info This function allows saving a file containing the IXC board Information. This file must be sent to CEIA service department. System Reboot This function allows performing a Reboot of the IXC System.
  • Page 87: System Info

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 3.23.2.14 System Info This function displays some relevant information about the Web Server. 3.23.2.15 Logout This function allows logging out the session. Simply click on the menu “Logout”: the dialog box for the login of a new user appears.
  • Page 88 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Page left intentionally blank...
  • Page 89: Appendix

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS 4 APPENDIX 4.1 Parameters list – Alphabetical order AC ..........32 EPC ........... 60 NMP ........... 50 ACCT ......... 51 EPCD ......... 60 NMS ........... 51 ADMS ......... 24 EPT ..........72 NMV ...........
  • Page 90: Parameters List - Menu Order

    FI002K0021v1100UK – THS/21E and THS/21 Programming Manual APPENDIX 4.2 Parameters list – Menu order Administrator ............ 15 Ejection .............. 28 Create user............15 Ejection mode ......EM ....28/29 Modify user ............15 Ej. distance ........ ED ....30 Erase user ........
  • Page 91 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual PROGRAMMING PARAMETERS Motor config ............49 Configuration-adv ..........37 Nom.MotorVolt....NMV ....49 IXC presence ......IXC ....37 Nom.MotorCurr....NMC ....50 Backup conf.to USB ..........38 Nom.MotorPower ...
  • Page 92 FI002K0021v1100UK – THS/21E and THS/21 Programming Manual APPENDIX Diagnosis managem......... 56 MD Test .............. 63 Photocell fault ......PA ....56 FE test ..............63 Sync.max.activ ..... SPMA ....56 NFE test .............. 63 Chck.max.activ ..... CPMA ....56 SS test ..........

This manual is also suitable for:

Ths/21

Table of Contents