Baskets; Anti-Tip Brackets - Thermo Scientific TSX Series Installation And Operation Manual

Blood bank refrigerators
Table of Contents

Advertisement

7.6 Baskets

7.7 Anti-Tip Brackets

7.7.1 Installation Instructions
Thermo Scientific
Blood Bank Refrigerators have the option to order 6 wire baskets.
Baskets are shipped in a factory installed position that promotes uniform spacing
between the baskets. The basket slides are adjustable to higher and lower
positions in the cabinet. You can position these slides in the vertical slots which
are spaced at one-inch intervals. The baskets must be removed in order to adjust
the position of the slides. The procedure for removing, reinstalling, and changing
position of the baskets can be found in Section 7.5.1, Section 7.5.2, and Section
7.5.3.
The maximum basket capacity is 27 kg (60 lbs).
WARNING: The refrigerator must be secured by the anti-tip bracket
supplied. Unless properly installed, the refrigerator could tip when drawers
are loaded. Personal injury and damage to the equipment and contents may
result from the refrigerator tipping.
This refrigerator has been designed to meet all recognized industry tip standards
for all normal conditions when anti-tip bracket is installed and properly engaged.
Installation instructions are provided for wood and concrete floors. Any other
type of construction may require special installation techniques as deemed
necessary to provide adequate fastening of the anti-tip bracket to the floor. For
installation on floors other than wood and concrete, please contact technical
support.
The use of this bracket does not prevent the tipping of the refrigerator when not
properly installed.
Figure 8. Materials Supplied
Label
Description
1
Bracket
2
Bolts
3
Anchors
4
Instructions and Installation Template
Installation
TSX Blood Bank Refrigerators
13