XI
Troubleshooting & Repair Procedures
RP19 - FCV Test
1) Fill the C1000/T with liquid oxygen. Allow approximately
one hour for the system pressure to stabilize. Verify that the
PRV is venting and that the system pressure is 1.28 – 1.62
bar (18.5 – 23.5 psi).
2) Attach a calibrated flow meter to the oxygen outlet barbed
fitting.
3) At each flow setting, check the flow control valve.
Acceptable flow ranges are given below in Table 9 and 10.
NOTE: Data listed in Table 9 and 10 is based on an operating
pressure range of 1.41-1.66 bar (20.5-24 psi). If the flow meas-
urements are out of specification, check the pressure in the unit. A
combination of high or low pressure and the tolerance of the par-
ticular flow meter you are using can result in inaccurate readings.
Table 9 C1000 FCV
Nominal Setting (LPM)
0.12
0.25
0.50
0.75
1.00
1.50
2.00
2.50
3.00
3.50
4.00
5.00
6.00
Allowable Range (LPM)
0.02 - 0.22
0.08 - 0.42
0.33 - 0.67
0.58 - 0.92
0.83 - 1.17
1.18 - 1.82
1.61 - 2.43
2.08 - 2.97
2.55 - 3.51
2.92 - 4.04
3.43 - 4.62
4.33 - 5.77
5.14 - 6.92
Table 10 C1000T FCV
Nominal Setting (LPM)
0.50
1.00
1.50
2.00
3.00
4.00
5.00
6.00
8.00
10.00
15.00
Companion Portable Technical Service Manual • PN 20562220 Rev B
Allowable Range (LPM)
0.33 - 0.67
0.83 - 1.17
1.18 - 1.82
1.61 - 2.43
2.55 -3.51
3.43 - 4.62
4.33 - 5.77
5.14 -6.92
6.72 - 9.28
8.42 - 11.53
12.97 - 17.28
38