Servicing - KNF NF 60 Operating And Installation Instructions

Diaphragm liquid pump
Table of Contents

Advertisement

Diaphragm liquid pump NF 60, NFB 60
Information on procedure
Prior requirements
KNF Flodos AG BA_NF60_EN_05_069208
Translated from the Original Operating and Installation Instructions
8.

Servicing

8.1. Servicing schedule
Component
Pump
Pump head
Diaphragm, valve plate
and seals
Tab. 13
8.2. Cleaning
WARNING
8.2.1. Flushing the pump
➔ If pumping aggressive media, KNF recommends flushing the
pump with air under atmospheric conditions for several
minutes prior to switch off (if necessary for safety reasons: use
an inert gas). This will extend the service life of the diaphragm.
8.2.2. Cleaning the pump
➔ Where possible, wipe the components with a soft dry cloth. Do
no use cleaning solvents as these may corrode plastic parts.
➔ If there is compressed air available, blow off the separate
parts.
Pump must be switched off and mains plug removed from
the socket.
The pump must be free of any hazardous substances.
Hoses must be disconnected from the pump head.
We recommend replacing the diaphragm when the head
parts are removed.
Tools
Qty.
Tool
1
Torxplus 10 IP screwdriver
Tab. 14
Servicing interval
-
Regular inspection for external
damage or leaks
-
Clean if the flow rate decreases, the
pump does not work or no vacuum
is created (Section 8.2)
-
Change as soon as pumping capac-
ity decreases, preferably sooner
Health hazard due to dangerous substances
in the pump!
Depending on the substance transferred, risk of
caustic burns or poisoning.
➔ Wear protective clothing if necessary, e.g. pro-
tective gloves.
➔ Rinse the pump with a neutral liquid and pump
empty.
Servicing
20