Safety Valve; Disassembly; Re-Assembly - Fluke P3000 Series User Manual

Pneumatic deadweight tester
Hide thumbs Also See for P3000 Series:
Table of Contents

Advertisement

Item

Safety Valve

Disassembly

Re-Assembly

1.800.561.8187
1
3
11
Figure 5-6. System Selector Valve
Table 5-6. System Selector Valve - Parts List
Description
1
Set Screw
2
Knob
3
Valve Shaft
4
Pin
5
Bearing
6
Valve Body
7
O-Ring
1. Unscrew safety plug (3) from the vacuum body (7), taking care not to lose the
bonded seal (6).
2. Carefully hold the larger "head" of the safety valve (1), and remove screw (5) and
valve stop (4).
3. Withdraw safety valve (1) from safety plug (3), and inspect O ring (2) for cleanliness
or signs of damage—replace as necessary.
Reassembly is the reverse of the above disassembly instructions, taking care that all
sealing surfaces are clean and undamaged.
www.
2
13
4
4
5
7
10
Part
Item
3927102
8
3927125
9
3921842
10
3906008
11
3905259
12
3921839
13
3905713
14
.com
Maintenance & Servicing
14
12
6
8
5
9
Description
Part
Valve Rotor
3921856
Screw
3908867
Bonded Seal
3921935
Connector
3868081
Support Bracket
3921492
Screw
2687927
Spring Plunger
3921874
information@itm.com
5
Safety Valve
gjn36.eps
5-13

Advertisement

Table of Contents
loading

Table of Contents