Introduction (cont.)
Screw/Pin Information
The chart below contains relevant information on various 3.2 mm
screws/pins that are compatible with the Persona System. If these
screws/pins are used during the procedure for instrument fixation,
they should be removed prior to closure as they are NOT implantable.
Screw/Pin
Screw/Pin Item #
25 mm x 2.5 mm
Female Hex Screw
42-5099-025-25*
75 mm x 3.2 mm
Trocar Tipped Drill
Pin (2.5 mm hex)
00-5901-020-00
Hex Headed Screw
33 mm long
00-5901-035-33
MIS Quad-Sparing™
Total Knee Headed
Screw 48 mm long
00-5983-040-48
25 mm Shorthead
Holding Pin
00-5977-056-03
* The 2.5 mm female hex screws and 2.5 mm male hex driver should not be used in cortical bone,
as this may increase the incidence of stripping of the driver.
Compatible Driver
2.5 mm Male Hex Driver
42-5099-025-00
Pin/Screw Inserter
00-5901-021-00
Pin/Screw Inserter
00-5901-021-00
Screw Inserter/Extractor
00-5983-049-00
Multi Pin Puller
00-5901-022-00
Persona The Personalized Knee Surgical Technique
5 |
Shipped Sterile/
Quantity
Non-sterile
per Package
Sterile
2
Sterile
4
Sterile
2
Sterile
1
Non-Sterile
1
Single use?
Yes
Yes
Yes
Yes
No
Need help?
Do you have a question about the Persona and is the answer not in the manual?