6 Operation
6.2 Preparing the sample
6.2.2 Preparing a sample for phosphor imaging
Required tools and materials
110
NOTICE
Do not scratch or bend the storage phosphor screen.
Keep the storage phosphor screen free from dust.
NOTICE
Protect the storage phosphor screen from water and volatile sol-
vents.
Dry the sample thoroughly before exposing it.
Wrap wet samples in plastic film and make sure that no liquid
exits. Use double layers if the sample contains volatile solvents.
If water enters the storage phosphor screen, the sensitivity of
the storage phosphor screen is reduced.
Volatile solvents may deform the protective film on the storage
phosphor screen.
NOTICE
Protect exposed storage phosphor screens from light until the
scanning is finished.
Storage phosphor screen cassette
Radioisotope-labeled samples
Gloves