Preparing A Sample For Phosphor Imaging - GE Amersham Typhoon Operating Instructions Manual

Table of Contents

Advertisement

6.2.2

Preparing a sample for phosphor imaging

Introduction
This section describes how to expose a storage phosphor screen to radioisotope samples,
and how to perform a scan of the storage phosphor screen.
Precautions
Amersham Typhoon Operating Instructions 29193226 AC
CAUTION
Always wear cotton gloves when handling storage phosphor
screens.
NOTICE
Avoid environmental radiation. Avoid exposing storage phosphor
screens in places where the environmental radiation may be in-
creased, for example rooms with concrete walls or in basements.
NOTICE
Remove the storage phosphor screen from the Amersham Typhoon
scanner when not in use. Store the storage phosphor screens in a
horizontal position, protected from moisture.
NOTICE
Do not stack cassettes during exposure.
NOTICE
Use a suction rod when taking the storage phosphor screen out
of the cassette. Prying the Storage phosphor out may result in the
edge peeling off, making the storage phosphor screen unusable.
6.2 Preparing the sample
6.2.2 Preparing a sample for phosphor imaging
6 Operation
109