Download Print this page

Lakeshore DRC-93C User Manual page 74

Temperature controller

Advertisement

Section IV
Model
DRC-93C
100 DISP
! Space
a
Line
110 O U T P U T 712;"W1"
! A and
B Input information
120 ENTER 712;A$
! Ask
for s t r i n g
W 1
130 DISP "W1 = ";A$
! Display
s t r i n g
W 1
140 DISP
! Space
a Line
150 OUTPUT 712;"W2"
! I n t e r f a c e S t a t u s
160 ENTER 712;A$
! Ask
for s t r i n g
W 2
170 DISP "W2 = ";A$
!
Display s t r i n g
W 2
180 DISP
! Space
a Line
190 OUTPUT 712;"W3"
!
Control D a t a
(Gain, R e s e t ,
etc.)
200 ENTER 712;A$
! Ask
for s t r i n g
W 3
210 DISP "W3 = ";A$
! Display
s t r i n g
W 3
220 DISP "Gain =";A$[1,3]
!
Display
Gain s e t t i n g
230 DISP "Rate =";A$[5,7]
! Display R a t e s e t t i n g
2 4 0
DISP "Reset =";A$[9,11]
! Display R e s e t s e t t i n g
250 DISP
" H e a t e r R a n g e
=";A$[13]
!
Heater
Range
260 DISP "% Power =";A$[15,17]
! % Power
270 DISP
! Space
a
Line
280 OUTPUT 712;"WS"
! S e t f o r W S
290 ENTER 712;A$
! Ask f o r
s t r i n g
W S
300
DISP " W S
=
";A$
! Display Sensor Reading
310 DISP
! Space
a Line
320 OUTPUT 712;"WC"
! S e t f o r W C
330
ENTER 712;A$
! Ask
for s t r i n g
W C
340
DISP "WC
=
";A$
!
Display Control
Sensor Reading
350 DISP
! Space
a
Line
360 OUTPUT 712;"WP"
! S e t f o r
W P
370 ENTER 712;A$
! Ask f o r
set
p o i n t d a t a
380 DISP "WP = ";A$
!
Display
s t r i n g WP
390 DISP
! Space
a Line
400
O U T P U T 712;"WY"
! S e t
for
WY
410 ENTER 712;A$
! Ask
for
Scan
Information
420 DISP "WY = ";A$
!
Display
Scan
Information
430 DISP
! Space
a
Line
440 OUTPUT 712;"WI"
! S e t f o r W I
450 ENTER 712;A$
! Ask
Input
Cards and Options
460 DISP "WI
=
";A$
! Display s t r i n g W I
470 END
4 . 1 6
SENSOR CURVE
PROGRAMMING
CONTROLLER
t e l l s
t h e DRC-93C t h a t
INSTRUCTIONS
when
it is
asked
t o
output d a t a ,
t h a t data
should be
the
output of
The
commands
which
w i l l
either
the standard Sensor Curves s t o r e d ,
output,
input,
e d i t
o r erase a
Precision
Option Curves s t o r e d and
Sensor Curve
are given i n
T a b l e
the
format
associated
w i t h
t h e
4-16.
I n
a d d i t i o n ,
the
commands
t o
REMOTE
S E N S O R I D Remote Position t o
assign
o r
change
assignments of the
Sensor
Curve assignments as
given
various curves t o
t h e
Sensor I D
i n
T a b l e
4-17.
This
output
is
t a b l e s (both A and
B) are given
i n
defined
as
the
Sensor Curve Infor-
Table 4-16.
mation
T a b l e
(SCIT)
.
As can
be
seen
from
the output
shown
on
this
4 . 1 6 . 1
The XDT
Command
page,
t h e
instrument
is
shipped
with
a l l remote
p o s i t i o n s c a l l i n g
T h i s
command
f r o m
t h e
BUS
up
Standard Curve 02.