Model
DRc-93C
Section
IV
display temperature is
123.45K
and
that the set point is
123.40K.
T a b l e 4 - 1 5 .
DRC-93C
Output Data Statements
Request
Output of Instrument Data
ws
wc
W P
wo
WM
Sample Sensor Data
8
Characters plus up to
2
Terminators where the
variations are the same as for WO (see below).
Control Sensor Data
8
Characters plus up to
2
terminators where the
variations are the same as for WO (see below).
Set Point Data
8
Characters plus up to
2
terminators where the
variations are the same as for WO (see below).
Sample (WS), Control Sensor (WC)
and
Set Point (WP) Data
26
characters plus up to
2
terminators where:
Display Math Data
57
characters plus up to
2
terminators where: