Technical Characteristics - OT Bioelettronica EMG-USB2 User Manual

Multichannel bioelectrical signal amplifier
Table of Contents

Advertisement

11. TECHNICAL CHARACTERISTICS

Model:
Risk class:
Insulation class:
Classification:
Case:
Power supply:
Consumption:
Limitations:
Working conditions:
Protections:
Input channels:
Amplifier:
Visualization:
Commands:
Dimensions:
Weight:
EMG-USB2
IIa in compliance with the standard 93/42/CEE.
BF type applied part, in compliance with the European standard EN 60601-1.
- class I, about the protection from indirect contact.
- IP20, about the penetration of fluids and dust; device not protected.
painted metallic case.
voltage from 90Vac to 260Vac ± 10%, frequency from 47 to 400Hz.
30 W.
the device is not suitable for use in environments with high oxygen
concentration and/or flammable fluids and/or gases; do not use with
electro-surgery or short wave/microwave therapy equipment.
device suitable for continuative work.
2 x 1,6A fuses.
up to 272 independent
Maximum input range:
Bandwidth:
Total noise (RTI)
Selectable gain:
Input impedance
CMRR
graphic LCD 128x64 pixel display.
7 keys protected by polycarbonate membrane.
350 x 271 x 120 mm
3 Kg.
50 mV
PP
3 ÷ 4400 Hz
< 4V
RMS
< 1V
RMS
100, 200, 500,
1000, 2000, 5000,
10000
> 10
11
 on the entire bandwidth
> 95 dB
pag. 29/31
EMG-USB2 User manual
(monopolar)
(differential)
v.2.0