CONNECTIONS SETUP
Before performing any measure it is necessary to set up all the instrumentation; this operation has
to be done keeping the equipment turned off. Be careful to connect all the cables properly to the
respective plugs and sockets.
To set up correctly the electromyograph follow the instructions hereinafter (FIG. 5):
Make sure that the PC used with the electromyograph is turned off.
Make sure that the EMG-USB2 power on switch is set in "O" position.
Connect the USB port, placed on the EMG-USB2 rear panel, to one of the PC USB port by
means of an A-B USB cable.
Connect the EMG-USB2 power supply connector, placed on the rear panel, to a 90÷260 V
50÷60 Hz power line supply voltage using the provided cable.
DANGER: the use of extension lead, multiple sockets or adapters can deteriorate the device
performances. Connection to sockets lacking in the ground conductor ("earth" conductor) or with a bad
quality of this connection can deteriorate the device performances and cause a potential risk for patient
and operator.
Turn on the PC (desktop computer or notebook).
Install the OT BioLab software following the instructions reported on the respective user
manual.
Computer with a free
USB port
Power line
socket
Power supply
cable
FIG. 5: Standard set up of the electromyograph