v.2.0

REAR PANEL

Figure 3 shows the connectors on the rear panel of EMG-USB2 described in the following sections.

Power Supply Switch

Fan
AUXILIARY INPUTS
AUXILIARY INPUTS
IN 1
IN2
IN 3
IN 4
IN 5
IN 6
IN 7
IN 8
IN 9
IN 10
IN 11
IN 12
IN 13
IN 14
IN 15
IN 16
0 1
USB
TRIGGER
ANALOG
OUTPUT
MADE IN ITALY
Fuse Box

Power Supply Socket

USB
TRIGGER
ANALOG OUTPUT
Connector
Connector
Connector
FIG. 3: EMG-USB2 rear panel view
Power supply socket
The EMG-USB2 must be connected to the power line socket only with the supplied cable. Ensure
that the wall socket is properly grounded.
DANGER: the use of extension cords, multiple sockets or adapters can impair the performance of the
EMG-USB2. Connection to sockets without proper grounding (e.g. lacking the "earth" conductor) or with
bad quality grounding can impair the performance of the EMG-USB2 and cause a potential risk for the
patient and the operator.
Power Supply Switch
The Power Supply Switch turns on/off the EMG-USB2. The switch position I turns the EMG-USB2
on; the switch position O turns it off.
The Power Supply Switch breaks both the power line wires to improve the safety. When the
EMG-USB2 amplifier is not in use, turn it off by this switch.