SECTION 3. GENERATOR PROBE & ACCESSORY MAINTENANCE

3.3 REASSEMBLY

1. Stand the new BEARING flat on a table surface and press the smaller
end of the ROTOR SHAFT through the BEARING.
2. Holding the ROTOR SHAFT off the table, push the ROTOR SHAFT
through the hole in the center of the collet wrench and use the wrench
to press the BEARING down the shaft until it snaps into place (the
bearing should feel loose in the bearing seat). (see figures)
4. Slide the COLLET NUT onto the ROTOR SHAFT.
5. Slide the FLEXIBLE COLLET about one centimeter onto the ROTOR
SHAFT.
6. Check that the unit is disconnected from the power source.
7. Insert the ROTOR SHAFT into the MOTOR unit until it stops. Ensure
that the shaft is all the way in the motor unit.
8. Slide the COLLET NUT down the ROTOR SHAFT and thread it onto
the MOTOR unit.
9. Hold down the SHAFT LOCK BUTTON and hand tighten the collet
nut as much as possible. DO NOT use the collet wrench.
10. Slide the GENERATOR TUBE over the ROTOR SHAFT and thread it
onto the MOTOR unit. Firmly hand tighten the tube.
11. See section 3.4 for instructions to run-in the new BEARING. (if the
bearing was not replaced, the Tissue Master is ready for use)
8