DENTSPLY X-SMART Endo Motor Operation Manual page 30

Table of Contents

Advertisement

X-SMART Endo Motor handpiece
Problem
The motor handpiece
does not rotate.
The motor handpiece
does not rotate. (The
E - 1
displayed.)
The motor handpiece
does not rotate.
tion speed are displayed
alternately.)
When turning on the
power, the alarm sounds,
and the motor handpiece
does not rotate.
The motor handpiece
keeps rotating. (The foot
pedal (Optional) does not
react.)
The motor handpiece
keeps rotating.
30
Cause
The motor handpiece cord is not
connected.
The foot pedal (Optional) is not
connected.
There is a break in the motor handpiece
or the motor handpiece cord.
The contra-angle head is clogged.
There is a short circuit inside the motor
handpiece or the motor handpiece
cord.
The contra-angle head is clogged.
-
There is a short circuit inside the motor
handpiece or the motor handpiece
cord.
The power is turned on while pressing
OFF button.
The power is turned on while pressing
the foot pedal (Optional).
There is a short circuit inside the foot
pedal.
It is rotated by the foot pedal
(Optional).
Check the connection.
Check the connection.
Contact your Technical Service
Department.
Clean or replace the contra-angle
head.
Contact your Technical Service
Department.
Clean or replace the contra-angle
head.
Contact your Technical Service
Department.
Contact your Technical Service
Department.
Check the foot pedal
(Optional).
tate the motor handpiece by the
dealer.
Stop the rotation by the
pedal (Optional) to stop, or pull
out the plug of the foot pedal
(Optional).
-