Page 2
They cannot be reproduced, translated, published or distributed without the permission of the copyright owner. Every effort has been made to ensure that all the information contained in the ADAMII™ User Manual is correct at the time of printing. However, NanoEntek Inc. reserves the right to make any changes necessary without notice as part of ongoing product development.
Introduction ADAMII is a bench top fluorescent cell counting analyzer that can be used in either a hospital and research laboratory. It has 4-channel (Bright, Yellow (PE), Red (PerCP) and Green (FITC)) light sources, so three color fluorescence samples can be analyzed simultaneously. The automatic X-Y-Z stage captures a maxi- mum of 136 images and which increases the accuracy of analysis.
Product components ADAMII™ is shipped with the following components. Once you receive your instrument, please check that all components listed below were shipped. If any component is missing or damaged, contact your local distributor or e-mail ivdst@nanoentek.com Power cord with 4 adaptor cords USB cable ADAMII™...
Product description Front view Insert/Eject Button Run Button State LED Insert/Eject Insert and eject the slide from the ADAMII™. button Functions as load and unload. Perform all procedures of automatic counting. Run button State LED Light up when software runs in the laptop and connection of camera and serial perform cor- rectly.
3. Place the instrument on a flat, level surface. 1. Plug the power cord into the electrical outlet and connect two USB cables.. Install the instrument Serial Camera Power supply port USB port 2. Connect the laptop and the ADAMII™ using two USB cables.
Page 9
Installation Install 1. Turn on the laptop and ADAMII™ by pressing the Power button. At this time, the slide holder will be ejected automatically. SW program 2. Double-click ADAMII™ icon on Desktop to start the software. ☞ Note: The ADAMII™ will be ready to use after initialization is completed when powered on.
Home menu Log in At the ‘Home’ screen, enter the user ID and password to log in. Default ID and password are both ‘master’ . User Click ‘User management’ when you need to manage accounts (e.g. ID addition, password change). To create a new ID, fill out the blank (user management name, user ID, password, confirm PW) and click ‘...
Page 14
[Attach dry-proof closure tape] Warning: • Calibration bead loading error - Ensure that 25μL of calibration bead is fully loaded on the ADAMII™ assay slide. - The instrument will not detect in low or high loading volumes. CAUTION: • Avoid bubbles as they may negatively affect results.
Page 16
Calibration ☞ Note: Please check ‘ADAMII™ Focus guide’ before setting the focus. Calibration procedure [Focusing example - representative images] Bright- Good Focus Bright- Bad Focus FL- Good Focus FL- Bad Focus • Bright - Good Focus: Clear circumference of the beads, Dark and clear contrast to the background •...
Page 18
Calibration Calibration 15. When the calibration is completed, the results will be displayed. procedure ☞ Note: If the results are not within the range on the tube, please ask your manufacturer or distributor. 16. When you need previous results of calibration, click ‘Record’ button.
Prepare sample and check the sample type. preparation To run a sample, click ‘Measure’ menu. Load sample Load 25 μL sample into the ADAMII™ assay slide and attach a dry- proof closure tape. [Load 25 μL sample] [Attach dry-proof closure tape] ☞...
Operation Insert 1. Open the ADAMII™ door. assay slide 2. Press to open up the slide holder. 3. Insert the assay slide into the slide holder. ☞ Note: Assay slide should be inserted to the end of the slide holder, in the direction of arrow on the slide, facing up.
Page 22
Warning: • Using ADAMII™ assay slide - Do not reuse the ADAMII™ assay slide, as leftover dye from the previous reading may affect the next reading. - Do not use any other slides as it results in inaccurate count and...
Operation To do a sample test, click the ‘Measure’ menu. Measure menu - Initial mode ☞ Note: Above screen and upcoming screen images only relate to CD34 application. 1. Counting mode 2. Display area 3. Detection channel 4. Insert sample 5.
Page 26
Operation Counting results of the sample will be displayed. It will be activated Measure menu after inserting sample. (Refer to page 33) - initial mode ⑤ Result area Date, sample/patient ID, and operator information will be displayed. ⑥ Test information...
Page 28
Operation Display preview image of the sample from the camera. Measure menu - Focusing mode ① Display area Indicate which channel is displayed. For ADAMII™-CD34 assay, ② Detection three fluorescent channels correspond to the markers below: channel • PerCP: CD45 •...
Page 30
Operation ☞ Note: Please check ‘ADAMII™ Focus guide’ before setting the focus. Measure menu - Focusing mode [Focusing example – bright field] ③ Focus option Bright- Good Focus Bright- Bad Focus #1 Bright- Bad Focus #2 Bright- Bad Focus #3 •...
Page 31
Operation ☞ Note: Please check ‘ADAMII™ Focus guide’ before setting the focus. Measure menu - Focusing mode [Focusing example – fluorescence channel] ③ Focus option CD45 PerCP- Good Focus CD45 PerCP- Bad Focus CD34 PE- Good Focus Bright- Bad Focus #2...
Page 32
Operation 1. Once the focusing is done, dilution factor and number of frames to Measure menu capture. If you want to see the viability, please check the circle - Focusing mode under the viability circle. ⑤ Run sample 2. Press the “Run sample” button. 3.
Operation Data menu provides the raw data including counting results, images Data menu from each frame in each channel. Users are able to review, save and export the data. Data list ☞ Note: Above screen and upcoming screen images only relate to CD34 application.
Page 36
• Purple: cells counted in PerCP channel (CD45 in ADAMII™-CD34 assay) • Green: cells counted in PE channel (CD34 in ADAMII™-CD34 assay) • Red: cells counted in FITC channel (dead cells in ADAMII™-CD34 assay) ☞ Note: This function is available only in the data menu.
Page 38
④ Sample volume The default numerical values of total and sample volumes for each sample type in ADAMII™-CD34 assay are shown below: information • HPC-A: total volume = 275 μL, sample volume 20 μL • Mobilized PB: total volume = 60 μL, sample volume 20 μL •...
Page 40
Operation How to find the invalid frames is described below: Data menu ⑦ Raw data table 1) Sample aggregation: when there is aggregation in specific area of sample, that frame should be deleted. Br image FL image 2) Dust: when there is dust inside the slide, that frame should be deleted.
Page 42
Operation To review the data that is saved in ADAMII™, select the desired sample from Data menu the data list. Scroll up and down to check the data of sample. ⑧ Data list Data list ADAMII™ allows you to export the data table(.csv), images(.jpg), ⑨...
Operation Setting menu contains information of ADAMII™ as well as several Setting menu user preferences and other Settings. It allows users to edit settings about the basic operation, software and firmware information. 1. Date & Time 2. Counting settings 3. Calibration 4.
Page 46
Operation Please refer to page 12~19 for calibration. Setting menu ③ Calibration To open the calibration result list, click ‘Record’ button. You can check the data of calibration briefly including date, time, lot number, user ID and result of each channel. Select the sample type.
Page 48
Operation [Firmware update] Setting menu 1. Save the firmware on USB disk and connect to the laptop. ⑥ Update 2. Press ‘FW update’ button. 3. Move to the folder where ‘Firmware update file’ is located and click the file. 4. Press OK button when the automatic update is completed. 5.
Cleaning and Maintenance Clean the surface of ADAMII™ instrument with a damp cloth. If liquid spills on ADAMII™, turn off the power immediately and wipe dry. ADAMII™ does not need regular maintenance. To troubleshoot problems with ADAMII™, contact technical support.
• Set the focus with 2ch manual focus function with autofocus function • First, check the sample type • Second, if you used the bottom of ADAMII™ CD34 Reagent Background of image is that almost used, please use the new reagent suddenly brighten •...
Page 52
• If none of the above 3 steps solve the problem, please request technical support • ADAMII™ does not insert and count the slide when the slide holder cap opened. Sensor of slide • But if sensor of slide holder cap is faulty and you start counting without holder cap is faulty closing the cap, then the result will not be accurate.
Page 54
Error code Error code Reason Message Position Data storage E0040 Data save is full Sample density Measuring E0050 is too high E0060 Update fail SW update...
Warranty NanoEnTek provides 1-year warranty service for defects of material and workmanship. If any defects occur in ADAMII™, NanoEnTek provides repair services for the defective parts at its discretion. The following defects, however, are specifically excluded: 1. Defects caused by improper operation.
Safety precautions Review and follow the safety instructions below: • If water or other material enters the instrument, adaptor, or power inlet, disconnect the power cord and contact a service person. For operating environment, refer to Product Specifications. • Do not touch the main plug or power cord with wet hands. •...
4. For more detailed information about disposal of your old appliance, please contact your city office, waste disposal service or visit our web-site, www.nanoentek.com. This product conforms to UL 61010-1, CAN/CSA C22.2 No.61010-1 “Safety Requirements for Electrical Equipment for Measurement, Con- trol, and Laboratory Use, Part 1: General Requirements.
Technical support Visit the our Website at www.nanoentek.com for : · Technical resources, including manuals, FAQs, etc. · Technical support contact information · Additional product information and special offers For more information or technical assistance, please call or email. Email ivdst@nanoentek.com...
Page 64
851-14, Seohae-ro, Paltan-myeon, Hwaseong-si, Gyeonggi-do, 18531, Korea Tel : +82-2-6220-7942, Fax : +82-2-6220-7999 MT Promedt Consulting GmbH Altenhofstrasse 80, 66386 St. Ingbert, Germany US Branch NanoEnTek USA Inc. 5627 Stoneridge Drive Suite 304, Pleasanton, CA 94588, USA Tel.: +1-925-225-0108, +1-888-988-0108(Toll free), Fax.: +1-925-225-0109...