Alarm Indicators
When an alarm occurs, the oximeter will indicate it through the
following indications:
Character flash
Alarm tone
High level alarms: character flashes quickly and sounds triple +
double + triple +double beep;
Medium level alarms: character flashes slowly and sounds triple
beep;
The sound pressure range for auditory alarm signal is from 45dB
to 85dB.

5.2 Alarm Conditions

5.2.1 Alarm off Before the First Measurement

Before the first measurement, the alarm system is configured to be
off. At this time, if the SpO
off, the oximeter will not give an alarm.

5.2.2 Alarm for SpO2 Sensor Unconnected

When the SpO
Medium alarm. The icon
SpO
and PR value area display "---", and give a medium alarm.
2
(Make sure the alarm system in menu is ON.)

5.2.3 Alarm for SpO2 Sensor off

When the SpO
give a medium alarm, and the icon
information area.
SpO
and PR value area display "---", and give a medium alarm.
2
(Make sure the alarm system in menu is ON.)

5.2.4 Alarm for Low Battery

sensor is unconnected or the sensor is
2
sensor is disconnected, the oximeter gives a
2
sensor falls off from the finger, the oximeter will
2
Unconnected
SpO
Low Battery
displays in information area.
42
Sensor off
2
displays in