Powered By Battery - Henry Schein HS-Vet Instructions For Use Manual

Vet pulse oximeter compatible with nellcor veterinary sensor
Table of Contents

Advertisement

Available languages

Available languages

The cable for connecting the oximeter and PC with the Oximeter
Viewer Data Management Software is also connected to this port.
Type BF applied part
Auxiliary output connector
SIO definition:
PIN
Name
1
RSGND
2
LED+
3
LED-
4
RXD
5
Detector Anode
6
Connection
7
AGND
8
TXD
9
Detector Cathode

2.4 Powered by Battery

The oximeter can be powered by four 1.5V LR6 AA alkaline
batteries. It will operate for 48 hours when used for general
operation, or about 24 hours of operation with the backlight and
alarm on.
The oximeter can also be powered by the Ni-MH rechargeable
battery package.
Battery Installation
To install the alkaline batteries:
1.
Make sure the oximeter is turned off.
2.
Press the battery compartment latch and remove the
battery access door.
Description
The RS232 GND
LED drive signal, IR Anode
LED drive signal, Red Anode
H100 RS232 RX
Detector anode
Detector connection
Analog GND
H100 RS232 TX
Detector cathode
21