The front panel of DI-4718B instruments provide screw terminal connection ports for analog input signal leads. The
front panel also provides remote events and stop/start remote recording control features via a built-in remote control
jack.

Signal Connections

The two 16-port screw terminals on the front of the instrument are used to interface analog input channels 1 through
8. Each channel has four terminals: In+, In-, Ex+, and Ex- for input signals and excitation (if required). Use the
sticker located on the top of your instrument for quick reference to terminal access port designations.
!
To avoid ESD damage in handling the device, take the following precautions:
Ground yourself with a grounding strap or by touching a grounded object before and during your handling of the
CAUTION
Never touch exposed screw terminal connector pins or screws.
instrument.