5.5.2 How to recharge the sensor

The power plug is designed to be used within 110-240V AC, 50-60Hz. Check if the required
voltage is available before connecting the adapter to the socket. Failure to do so can lead to
damage to the sensor, charging cradle, power plug, or charging cable.
The sensor can only be charged using the supplied wireless charging cradle.
To bring a depleted battery up to 10 hours of usage requires approximately 60 minutes of charging. A
complete recharge requires approximately 4 hours.
To charge the sensor:
1. Find a power source and connect the power plug and charging cable to it.
2. Connect the charging cable to the charging cradle.
3. Position the sensor into the charging cradle with its green side up such that the sensor fits in the
charging cradle's cavity. A blue indicator light will appear on the charging cradle to indicate that
the sensor is charging.
4. The sensor is fully charged when the blue indicator light on the charging cradle switches off. Make
sure this is not due a disconnection with the electricity source.
It is safe to leave the sensor on the charging cradle even when charging is complete.

6 Care and maintenance

Regular cleaning and maintenance should be carried out on the NightOwl® Sensor as described below.

6.1 Cleaning

 Never use abrasive agents, chlorine-containing substances, acetone, or other solvents to clean
the device.
 Do not immerse the sensor or charging cradle and ensure that no fluids penetrate the products.
 Do not attempt to sterilize the device, as this could cause unobservable damage to the inside
of the unit.
1.
Turn off the sensor and ensure the charging cradle, charging cable, and power plug are
disconnected from any source of power.
2.
Clean the following parts with a damp cloth and a mild liquid soap: sensor and charging cradle.
Ensure that no liquid enters the USB-C port of the charging cradle.
3.
Leave the cleaned parts to dry.
Instructions for Use – NightOwl
®
Sensor – Healthcare Professional
Page 17 of 24