2.7 -Developer display temperature not the same as actual temperature.

Are there volatile changes in environment (: room temperature)?
If temperature in dark room is unstable, wait until the temperature is stabilized and recheck it.
Check the chemical circulation
In the event there are air bubbles in the hoses or pumps
section 1.5.
Chemical is not circulated because of contamination and clogging of Drain Filter
Drain Filter.

2.8- Over or under supply of water

Adjust the water supply.
Check the water pressure, reduce or increase pressure as needed.
Check the operation of the Film Detection Sensor.
REFER TO TROUBLE SHOOTING SECTION 1.2
Check the operation of the Water Level Sensor.
Make sure the level sensor behind the Wash Rack moves freely.
Remove any particles affecting the float and clean it to ensure the float moves
freely.
Check if the Level Sensor Bracket is securely in place. If the sensor is not securely
in place it will not function properly.
JP-33 SERVICE MANUAL
- -
23
- Trouble Shooting
Refer to trouble shooting
Clean