Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Tele P atch™ Cardiac Monitor PM750 User Manual INSTRUCTIONS FOR USE Tele P atch is intended for use as prescribed by a physician who wants to follow cardiac activity.
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Table of Contents GETTING STARTED AND INTRODUCTION Notices, Cautions, and Copyrights TelePatch Cardiac Monitor User Kit Contact Us ABOUT TELEPATCH CARDIAC MONITOR PM750 INDICATIONS FOR USE OVERVIEW SAFETY SPECIFICATIONS AND COMPLIANCE...
Page 3
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Step 4 - Turn on Pendant and Connect Step 5 - Starting the Handset Step 6 - Procedure Set Up Step 7 - Procedure Settings Pacemaker Settings Protocol Setting Set Up...
Page 4
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 EMERGENCY AND SUPPORT NUMBERS Call 911 PROCEDURE INFORMATION AND CAUTIONS TELEPATCH SYSTEM WHEN WEARING ELECTRODE PATCH SHOWERING OR SWIMMING WITH TELEPATCH SYSTEM AND ELECTRODE PATCH TELEPATCH SYSTEM WHEN USING CABLE CRADLE SHOWERING OR SWIMMING WITH TELEPATCH SYSTEM AND CABLE CRADLE / HANDSET...
Page 5
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Unable to find Pendant Pendant is Turned Off Procedure Has Ended Start Procedure Exceptions Selecting a Pendant SAFE BATTERY USE AND DISPOSAL Safely Loading the Pendant Battery Battery Maintenance Disposal of Rechargeable Lithium Ion Batteries EXPLANATION OF MARKINGS...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 TelePatch Cardiac Monitor User Kit The standard TelePatch Cardiac Monitor PM750 kit comes with: 1. U niversal Pendant 2. Electrode Patch with Cradle 3.
Page 8
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 TELE P ATCH KIT CONTENTS Item Quantity Picture Legal Booklet Tele P atch Pendant Electrode Patch *Kit will have one or the other patch style, single or dual is based on procedure prescribed by physician with Cradle...
(3-5 per pouch) 1- 4 Breakaway Lanyard Contact Us Contact Medicomp, Inc. for any issues concerning the T ele P atch product such as: • General questions about our product • Product safety • Safe disposal of component parts •...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 600 Atlantis Rd. Fax: 321-676-2282 Melbourne, FL, 32904 Website: www.medicompinc.com email: info@medicompinc.com ABOUT TELE P ATCH CARDIAC MONITOR PM750 INDICATIONS FOR USE The TelePatch Cardiac Monitor PM750, is a pager-sized, handheld or patient worn device designed specifically to record and transmit ambulatory ECG signals.
Atmospheric Pressure: Operating: 700 hPa to 1060 hPa; Storage/Transport: Not Applicable to T ele P atch System ● The equipment contains no user-serviceable parts. It shall be serviced only by Medicomp, Inc. Unauthorized repairs of the equipment will void the warranty.
THIS IS A PRESCRIBED MEDICAL DEVICE, NOT A TOY, INFANTS AND CHILDREN MUST BE SUPERVISED. ● WARNING: CHOKING HAZARD – ADULT SUPERVISION REQUIRED. ● IN THE EVENT OF A DAMAGED PENDANT, DISCONTINUE USE AND CALL MEDICOMP PATIENT SUPPORT: 800-234-3278 EXT. 2370, FOR RETURN AND REPLACEMENT. COMPLIANCE Safety Classification In accordance...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. This Pendant complies with Industry Canada license-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this Pendant may not cause interference, and (2) this Pendant must accept any interference, including interference that may cause undesired operation of the Pendant.
Page 14
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 The Pendant POWER Button This is the Pendant POWER button. It is located on the FACE of the Pendant. The Pendant Symptom Button This is the Pendant Symptom button.
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 3G Connectivity Icon Bluetooth Connectivity Icon The Handset Battery Strength Indicator This is the Handset Battery Strength Indicator in percentages. This is located in the lower left corner of the Handset screen once Bluetooth communication has been set up.
Page 16
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Below is a graphical overview of the T ele P atch System. Detailed description of icons, buttons, and screen indicators are in previous section. KEY DESCRIPTIONS PROVIDED BELOW ARE FOR USE WITH THE ...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 LG Handset Front of Tele P atch H andset Back of Tele P atch H andset PENDANT AND Handset PICTURED ARE NOT ACTUAL SIZE. TELE ...
Page 18
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 The Pendant SYMPTOM Button The U niversal Pendant has a Symptom* button which is located in the center of the face of the Pendant.
Page 19
U niversal Pendant . THE RECHARGEABLE BATTERY PROVIDED WITH THE TELEPATCH SYSTEM IS NOT TO BE DISCARDED IN HOUSEHOLD WASTE. IF A BATTERY GETS DAMAGED CONTACT MEDICOMP, INC.: 877-996-5553. Tele P atch™ C ardiac Monitor PM750 USER MANUAL - UTM0000701-05E ...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Battery Well Battery properly inserted in Battery Well Detailed directions to change and charge the battery properly are found on page 25 or page 68. IT IS VERY IMPORTANT TO TURN THE PENDANT OFF BEFORE CHANGING BATTERY, OTHERWISE DATA MAY BE LOST.
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 LG Handset Accessories to the Tele P atch System Tele P atch™ C ardiac Monitor PM750 USER MANUAL - UTM0000701-05E | APRIL 4, 2019...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Tele P atch Charger Cord The T ele P atch Charger Cord is compliant for use with the Handset and the Battery Charger. Both are components of the ...
Page 23
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 For optimal performance from the Handset, it is recommended to plug the charger cord into the Handset overnight until the Handset is fully charged. Instructions on charging the Handset are on page 22 of this User Manual.
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Connecting Charger Cord: Handset Motorola Handset The port for the Motorola Charger Cord is located on the left side of the Handset near the bottom Charger properly inserted Correct direction of charger head...
Page 25
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 LG Handset The port for the LG Charger Cord is located on the bottom side of the Handset. Charger properly inserted Correct direction of charger head Incorrect direction of charger head Connecting the Charger Cord to the Handset: Take the end of the Charger Cord and insert it into the lower left side or bottom of the Handset...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Handset IS POWERED BY THE CHARGER CORD ONCE CHARGER CORD IS PLUGGED IN TO BOTH Handset AND WALL OUTLET. CHARGER CORD MUST NOT BE OBSTRUCTED BY FURNITURE OR LARGE OBJECTS AND REMAIN ACCESSIBLE TO DISCONNECT FROM WALL OUTLET EASILY, AS NEEDED.
THIS PRODUCT UTILIZES A RECHARGEABLE LITHIUM-ION BATTERY PACK. TO ENSURE SAFE CHARGING AND MAXIMUM PERFORMANCE OVER THE LIFE OF THE PRODUCT ALWAYS USE THE MEDICOMP PROVIDED CHARGER. DO NOT ATTEMPT TO CHARGE THE BATTERY ON ANOTHER MANUFACTURER’S CHARGER. Tele P atch™ C ardiac Monitor PM750 USER MANUAL - UTM0000701-05E ...
Page 28
Call M edicomp ’ s Patient Support: 8 77-996-5553 ● Place a Battery from the T ele P atch K it into the Battery Charger, with the Medicomp label facing up and the two small tabs into the notches.
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Flashing Orange Light: Battery is charging Green Light: Battery is charged MAKE SURE WALL OUTLET IS NOT CONTROLLED BY A LIGHT SWITCH. BATTERY CHARGER IS POWERED BY THE CHARGER CORD ONCE CHARGER CORD IS PLUGGED IN TO BOTH BATTERY CHARGER AND WALL OUTLET.
Page 30
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Single Channel Electrode Patch with Cradle Dual Channel Electrode Patch with Cradle The Universal Pendant slides into the cradle on the Electrode Patch. Placing Pendant into Electrode Patch cradle: ●...
Page 31
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 ● Single Channel Electrode Patch supports modified lead II with reference right leg drive. ● Dual Channel Electrode Patch supports Channel A modified lead II, Channel B modified III and a reference electrode.
For TelePatch Wireless Single Channel 1. LEFT side of chest. 2. Top of patch (Medicomp logo right-side up) near top of breast bone. 3. Bottom of patch angled toward left nipple at a 45- to 60- degree angle. For TelePatch Wireless Dual Channel: 1.
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Dual Channel and Single Channel Electrode Patch Placement Step 2 - Prepare the Skin Because the Electrode Patch will be worn for up to seven (7) days, proper skin preparation is important for the procedure.
● Once it is firmly adhered to skin peel remaining white tabs on the patch so the whole surface of the patch is adhered to skin. THE MEDICOMP ECG ELECTRODE PATCH IS SINGLE USE, NON-STERILE, AND DISPOSABLE AND IS TO BE USED ON INTACT (UNINJURED) SKIN .
Hold Pendant face down in hand or on a flat surface. There is a well for Battery to be ● inserted Insert the Battery into the opening, with the Medicomp label facing up and the two ● small tabs into the notches, as pictured.
Page 36
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Slide the latch at the back of the Battery to the left, securing the Battery in place ● Battery is NOT secure in Pendant – Unlocked Battery IS secure in Pendant - Locked Turn the Pendant over, slide the Pendant into the Electrode Patch Cradle with the lock ●...
Page 37
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 ● For a Holter procedure: (24 hour to 96 hour) o Press the Symptom button three (3) times. ● For all other procedures: (7 day to 30 day) o The Pendant will initiate on its own.
● Take the Handset and turn it on by pressing and holding the POWER button which is located in the upper right side of the Handset. ● Wait for the Medicomp main screen to appear on the screen before continuing with standard procedure setup.
After following the instructions on the patient preparation screen, which may have already been completed, the next step is to select the correct settings before starting a procedure. Medicomp Tele P atch Handset Patient Preparation Screen ● Tap “OK”...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Medicomp TelePatch Handset Procedure Set Up Screen Choose “Accept Settings” or change settings from this screen. To accept default settings: ● Tap “Accept Settings”.
THE T ELE P ATCH SYSTEM CAN BE USED FOR INFANTS WEIGHING LESS THAN 10KG. “TEST”SHOULD NEVER BE SELECTED AS A SETTING WITHOUT THE SUPERVISION AND GUIDANCE OF A MEDICOMP CARDIAC MONITORING CENTER (CMC) TECHNICIAN. CALL THE CMC FOR GUIDANCE IN “TEST” MODE.
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Set Up All TelePatch systems will be shipped in “Normal” mode. There is therefore, no need to instruct users on how to select the “Normal” mode. The following instructions describe the steps and screens for selecting “Test”...
8 77-996-5553 Step 8 - Initial Procedure Screens After applying and accepting the procedure settings, the procedure will initialize. Medicomp Tele P atch Handset Scanning for Universal Pendant Medicomp Tele P atch Handset Medicomp Tele P atch Handset...
Page 44
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 The Medicomp TelePatch Handset Monitoring Screen The T ele P atch p rocedure is now running. ● IT IS VERY IMPORTANT TO TURN THE PENDANT OFF BEFORE CHANGING BATTERY, OTHERWISE DATA MAY BE LOST.
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Keep the Handset charged and turned on. ● Keep Handset within 10’ of the Pendant during procedure. ● Review how to record a SYMPTOM, make a Diary Entry. ●...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 How to Make a Diary Entry: It is important the prescribing physician to know when a symptom is felt and activity at the time the SYMPTOM button was pressed: While “Recording in Progress”...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Symptom Menu Location Menu Activity Menu Audio Diary Entry: Tap “Audio”. ● Clearly state the symptom, location ● where symptom occurred, and activity during symptom.
ECG strips show Lead II at the top, and Lead V1/MCL1 on the bottom. 3-Wire Hook-up: Lead II and Lead I are used. [Referring to the Medicomp reports, the ECG strips show Lead II at the top, and Lead I on the bottom...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 3-Wire Hook-up 5-Wire Hook-up For women with large breasts, the following electrode positions are recommended: ● Red in the left axillary line ●...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 The above skin prep procedure is only required at the beginning of the procedure, during Handset initialization. Users will not be required to perform this type of preparation each time they relocate electrodes.
Page 51
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 ● Snap the wires onto the electrodes while they are still attached to the plastic backing. ● One by one, peel each electrode from the backing and apply it to the chest. ●...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 PEOPLE WORKING WITH MACHINERY OR WORKING IN ENVIRONMENTS WHERE LOOSE HANGING ROPE- LIKE OBJECTS CAN POSE A POTENTIAL THREAT OR HARM TO THEMSELVES AND/OR THE MACHINERY ARE ADVISED TO CARRY CABLE CRADLE WITH PENDANT IN POCKET.
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 ● The Pendant should always be in the Cable Cradle, the Cable should remain plugged into the Cable Cradle. ● Enjoy all of your regular activities like exercise or gardening, while wearing the Pendant, and Cable Cradle with Cable.
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 ● Place index fingers firmly on the Fingertip Electrodes (the silver-gray corners) on the top edge of the Pendant for 30 seconds. ● Handset will display “Recording in Progress”, keep fingers on the Fingertip Electrode corners until the Handset screen message says “Recording Complete”...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Can I Shower? When wearing the U niversal Pendant with the Electrode Patch you can shower with the Pendant and patch on. When done in the shower gently pat (not rub) the face of the patch and the Pendant dry.
This is a good way to secure lead wires to prevent pulling, thus minimizing noise interference during the procedure. Wear the Cable Cradle with the lanyard or keep the Cable Cradle in a pocket. Note: Stress loops are not provided by Medicomp Inc. EMERGENCY AND SUPPORT NUMBERS Call 911 Below are instructions on how to call 911 from the Handset.
RETURN U AND ALL COMPONENTS IN KIT PROVIDED, TO NIVERSAL ENDANT MEDICOMP, INC., USING THE INCLUDED POSTAGE PAID ENVELOPE. SHOWERING OR SWIMMING WITH TELEPATCH SYSTEM AND ELECTRODE PATCH ● THE U IS WATER RESISTANT. YOU CAN SHOWER WITH THE NIVERSAL ENDANT AND ELECTRODE PATCH INTACT.
, AND ALL COMPONENTS IN KIT PROVIDED, TO NIVERSAL ENDANT MEDICOMP, INC., USING THE INCLUDED POSTAGE PAID ENVELOPE. SHOWERING OR SWIMMING WITH TELEPATCH SYSTEM AND CABLE CRADLE ● THE CABLE CRADLE SHOULD NOT BE WORN IN THE SHOWER. ● WHEN WEARING TELEPATCH WITH A CABLE CRADLE AND INDIVIDUAL...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 SKIN. IF AN INDIVIDUAL ELECTRODE SELF-REMOVES IN SHOWER, PREPARE THE AREAS AND APPLY NEW ELECTRODES. ● IF SHOWERING WITH ELECTRODES, GENTLY PAT THEM DRY WITH A TOWEL BEFORE RECONNECTING THE LEAD WIRES TO THE ELECTRODES ●...
Hz with respect to the system ground. ● This is a prescribed medical Pendant, not a toy; keep away from small children. ● In the event of a damaged Pendant, discontinue use and call Medicomp’s Patient Support: 8 77-996-5553...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Use a tissue, lightly moistened with alcohol or a hospital-grade cleanser such as Sporicidin – (Do not saturate) - wipe off debris, dirt or other matter that may gather on either the Handset or Pendant of the ...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 Symptom Recording “Recording Complete” message None Complete on Cables or Patch will be seen on screen of Handset Finger Tip Electrodes Start of Green LED 3 flashes 0.5 second Beep 3 times in unison with Recording...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 recording Early low main battery Orange LED blink 1 per Beep 1 second duration every warning second 4 minutes Late low main battery Orange LED blink 3 times per Beep 1 second duration every warning...
All TelePatch systems are shipped in the 'Setup/Normal' mode. 'TEST' SHOULD NEVER BE SELECTED AS A SETTING WITHOUT THE SUPERVISION AND GUIDANCE OF MEDICOMP PATIENT SUPPORT. THE PATIENT SHOULD CALL MEDICOMP FOR GUIDANCE IN THE 'TEST' MODE. During the Quality Hookup Tests, the Handset may display the following...
The results of the skin prep test may be displayed before the procedure begins. -• Go to PATIENT PREPARATION for additional steps. • For additional troubleshooting instructions, call Medicomp’s Patient Support: 877-996-5553. Error Displays During procedure hookup, the T ele P atch tests for flex circuit or cable breakage, and checks that its minimal requirements for skin prep and signal voltage are met.
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 If inadequate signal voltage (can be either low or high) in one channel while the other one is acceptable, place the electrodes in a “parallel”, or ”side-by-side” lead arrangement. Identify the channel that has the acceptable voltage and place the electrodes of the problem channel side-by-side with it.
Cables (lead wires) not connected If the electrode cables are disconnected, this error message will also appear. • Go to PATIENT PREPARATION for additional steps. • For additional troubleshooting instructions, call Medicomp’s Patient Support: 877-996-5553. High/Low Voltage Notifications LOW VOLTAGE IN CHANNEL A (WHITE [A-]/RED [A+] ELECTRODE PAIR) Although the ...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 • For additional troubleshooting instructions, call Medicomp’s Patient Support: 877-996-5553. LOW VOLTAGE IN CHANNEL B (BLACK [B-], RED [A+] ELECTRODE PAIR The Handset's message for a low voltage in Channel B is: Low B, Move Black...
Procedure Has Ended When the procedure ends, an alert will be displayed. Users should return the unit to the Medicomp address provided in the Patient Guide. Every T ele P atch Kit has a postage paid return envelope. Include all components in the zippered kit and mail from any mailbox.
The TelePatch Cardiac Monitor Pendant operates on a single, 3.7V lithium ion battery. Bottom and top views of the Universal Pendant Battery IF PENDANT’S BACKUP BATTERY IS LOW, PATIENT SHOULD CALL MEDICOMP PATIENT SUPPORT: 800-234-3278 ext. 2370. IT IS VERY IMPORTANT TO TURN THE PENDANT OFF BEFORE REMOVING AND LOADING BATTERY, OTHERWISE DATA MAY BE LOST.
● Hold Pendant face down in hand or place on a flat surface. There is a well where battery will be inserted ● Insert battery into well of Pendant: with the Medicomp label facing up, and the two small tabs into notches in flat wall of the well Inserting battery into Pendant battery well ●...
● The rechargeable battery provided with the TelePatch System is NOT to be discarded in household waste. ● If a battery gets damaged contact Medicomp, Inc.: 8 77-996-5553 EXPLANATION OF MARKINGS Equipment markings and caution labels are important for safe and reliable use; they must not be removed.
Page 73
The WEEE logo on this product indicates that this product Battery must not be disposed of with standard waste but rather recycled. Please contact Medicomp, Inc. for more information on proper component disposal. RoHS Compliance This product is in compliance with Directive 2002/95/ EC of...
Page 74
Catalog Number/Part Number Lithium Ion Item is recyclable. Battery Users should properly dispose of components through Medicomp, Inc. Please contact Medicomp, Inc. for information on how to return item for recycling. Lithium Ion EU regulation of chemicals and safe use: Registration, Battery...
● The Universal Pendant should NOT be treated as household waste. ● The Electrode Patch or individual electrodes ONLY should be discarded after use. ● Return TelePatch, Pendant, and all components in kit provided, to Medicomp, Inc., using the included postage paid envelope.
INFORMATION PROVIDED IN THIS MANUAL . U SE THE T ELE P ATCH SYSTEM ONLY WITH THE LEADS, ELECTRODES, AND ACCESSORIES RECOMMENDED BY MEDICOMP. USE OF OTHER ACCESSORIES MAY ADVERSELY AFFECT THE PERFORMANCE OF THE DEVICE OR MAY RESULT IN STRONGER ELECTROMAGNETIC EMISSIONS OR REDUCE THE ELECTROMAGNETIC IMMUNITY OF TELEPATCH CARDIAC MONITOR PM750...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 RF emissions CISPR 11 Class B The PM750 is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC...
Have a question or need assistance? Call M edicomp ’ s Patient Support: 8 77-996-5553 0deg 0% of Vnom, 1.0 cycles at 0deg 0deg 0% of Vnom cycles at 0deg 0% of Vnom cycles at 0deg 0% of Vnom cycles at 0deg Power frequency (50/60 Hz) magnetic...