Resolution - Endress+Hauser Levelflex FMP53 Technical Information

Guided level-radar, level measurement in liquids and hygienic applications
Hide thumbs Also See for Levelflex FMP53:
Table of Contents

Advertisement

Levelflex FMP53

Resolution

Endress+Hauser
Differing from this, the following measuring error is present in the vicinity of the lower probe
end:
1
80
60
40
20
0
-20
-40
-60
-80
0
50
100
150
A[mm]
1
Rod or coax probe
2
Rope probe
A
Distance from probe end
D
Sum of non-linearity, non-repeatability and hysteresis
If for rope probes the DC value is less than 7, then measurement is not possible in the area of the straining
weight (0 to 250 mm from end of probe; lower blocking distance).
Differing from this, the following measuring error is present in the vicinity of the upper probe
end (rod/rope only):
40 mm
30 mm
20 mm
10 mm
3 mm
D
0 mm
- 3 mm
- 10 mm
- 20 mm
- 30 mm
- 40 mm
D
Sum of non-linearity, non-repeatability and hysteresis
R
Reference point of measurement
DC Dielectric constant
• digital: 1 mm
• analog: 1 mA
2
DC = 2
DC = 80
200
250
300
R
100 mm
80
60
40
20
0
-20
-40
-60
-80
0
50
100
150
200
A[mm]
DC = 2
DC = 80
200 mm
DC = 80
250
300
A0016586