• ECG NEO MODE
Simulates a NSR neonatal patient at 140 BPM with:
• ART pressure on P1 of 60/30
PA pressure on P2 of 20/10
LV Pressure on P3 of 60/5
CVP Pressure on P4 of (2) mean
• 100 mmHg MODE
Simulates static 1 OOmmHg on all four channels.
• 200 mmHg MODE
Simulates static 200mmHg on all four channels.
• IBP STEPS MODE
In this mode, a selection of static blood pressures are available on all four channels. The
pressures available are shown on the bottom of the unit and are 0, 25, 50, 100, 150, 200, and
250mmHg. When entering the mode, the pressure will be OmmHg, so zero the monitor while in
this step. Click the MODE button to advance to the next pressure step.