Summary of Contents for Pronk Technologies SimSlim SL-8

  • Page 1 SimSlim Patient Simulator Operator's Manual PReNK TECHNOLOGIES www.pronktech.com ©2014...
  • Page 2 SimSlim, Pronk, and Pronk Technologies are trademarks of Pronk Technologies Inc. ©2012 Pronk Technologies Inc. All rights reserved. CAUTION: Risk of explosion if lithium battery is replaced by an incorrect type. Dispose of used batteries according to country and state specific laws and regulations.
  • Page 3: Table Of Contents

    Getting Started Detailed ECG Operation Detailed Invasive Blood Pressure Operation 8 Detailed Cardiac Output Operation Detailed YSI Temp Operation Troubleshooting Tips Services Warranty and Repair Information Speci cations Revision L 07/18/2017...
  • Page 4: Getting Started

    Introduction The S/mSl/m ,.., simulator is a compact, rugged, easy to use multi-parameter patient simulator with the following features: • 10 Snap/Banana plug ECG, Resp and Pacer Simulation. • 4 Invasive Blood Pressure Channels via four 6 pin mini-din connectors. •...
  • Page 5 3. Waveform Selection Within a Mode • Some of the SimSlim simulator modes contain multiple waveforms. • An example of this is ARR MODE 1 which has NSR with PVCs, VTACH, VFIB and Asystole waveforms. • These waveforms are listed on the bottom label. •...
  • Page 6: Detailed Ecg Operation

    Detailed ECG Operation • Connecting the ECG Leads ECG lead wires with snaps can be connected directly to the SimSlim simulator snaps. The snaps are clearly labeled with the name and color code of the lead. If your lead wires have banana plugs rather than snaps, the banana plugs are slid into the grooves under the snap such that the metal part of the plug is against the underside of the snap.
  • Page 7 MODE Revison L 07/18/2017...
  • Page 8: Detailed Invasive Blood Pressure Operation

    6 pin min-din simulator IBP pin-out is fairly standard. Wiring is as follows: + Excit = pin 1, - Excit = pin 4, + Sig = pin 3, -Sig = pin 6. Contact Pronk Technologies Inc. at 800/541-9802 for more information.
  • Page 9 • ECG NEO MODE Simulates a NSR neonatal patient at 140 BPM with: • ART pressure on P1 of 60/30 PA pressure on P2 of 20/10 LV Pressure on P3 of 60/5 CVP Pressure on P4 of (2) mean • 100 mmHg MODE Simulates static 1 OOmmHg on all four channels.
  • Page 10: Detailed Cardiac Output Operation

    Detailed Cardiac Output Operation • Connecting the Cables The SimSlim simulator is designed to operate without a custom C.O. cable. Instead, use the monitor's standard C.O. cable. The connections are as follows: 1. Connect the monitor's C.O. cable to the monitor. 2.
  • Page 11 Unfortunately, different monitor vendors use such a wide variety of thermistors and connectors for injectate temperature that Pronk Technologies cannot supply these. One solution would be to fabricate an injectate temp simulator locally from an injectate probe and some resistors. Some vendors use YSI 400 or YSI 700 thermistors for injectate temp.
  • Page 12: Detailed Ysi Temp Operation

    cable and a 3.5mm to 1/4”stereo plug converter. 1/4” monitors 1/4” Revision L 07/18/2017...
  • Page 13: Troubleshooting Tips

    (Prices subject to change without notice) SERVICE ITEM NUMBER DESCRIPTION PRICE (USO$) Annual SimSlim Calibration Full checkout, adjustment and Call for Pricing calibration certificate *Pronk Technologies recommends an annual calibration interval. PRC:,NK TEOHNC:,LC:,GIES (800) 541-9802 Page 13 Revision L 07/18/2017 www.pronktech.com...
  • Page 14: Warranty And Repair Information

    If any fault develops, notify Pronk Technologies (see Returns and Repairs, below) giving full details of the difficulty, and include the model and serial number of the device. Upon receipt of shipping instructions, forward the device prepaid and repairs will be made at the factory.
  • Page 15: Speci Cations

    PRc:>NK TEOHNc:>Lc:>C31ES www.pronktech.com (800) 541-9802 Page 15 Revision L 07/18/2017...