Detailed Invasive Blood Pressure Operation - Pronk Technologies SimSlim SL-8 Operator's Manual

Patient simulator
Table of Contents

Advertisement

Detailed Invasive Blood Pressure Operation

• Connecting the Cables
You will need a monitor specific cable to connect the SimSlim simulator IBP simulation ports to
your monitor.
A variety of these cables are available from Pronk Technologies and other
sources. Pronk's cables are supplied as a 6' (1.8 m) extension cable which connects to a 6"
(15cm) adapter cable. Most simulator cables with 6 pin mini-din connectors should work, as the
6 pin min-din simulator IBP pin-out is fairly standard. Wiring is as follows: + Excit = pin 1, - Excit
= pin 4, + Sig = pin 3, -Sig = pin 6. Contact Pronk Technologies Inc. at 800/541-9802 for more
information.
T he IBP simulation ports are labeled on the bottom of the unit.
• Zeroing Simulated Blood Pressures
In all modes except IBP STEPS, clicking the mode button will force all four IBP channels to
OmmHg for 15 seconds. If you desire to return to normal operation prior to the 15 second
timeout, a second MODE button click will 'unzero' all channels. (Note that this 'click to unzero
function' is not reliable in Beta software prior to B.3)
• Sensitivity
T he SimSlim simulator simulates 5uVN/mmHg transducers.
• ECG NSR MODE
Simulates a NSR adult patient at 60 BPM with:
• ART pressure on P1 of 120/80
• PA pressure on P2 of 30/1 O
LV Pressure on P3 of 125/15
• CVP Pressure on P4 of (6) mean
PRc:>NK T'EOHNc:>Lc:>C31ES
Revision L 07/18/2017