• Starting a Simulated Injection
Hold the MODE button down for 2 seconds to start a simulated injection. When the injection is
started, the LED will flash. If you are doing C.O. simulations for a long period of time, keep in
mind that the injectate temperature may drift and the C.C. value may have to be updated from
time to time for best accuracy.
• Using an lnjectate Temp Simulator
It is possible to use an injectate temp simulator rather than the actual injectate temp probe.
Unfortunately, different monitor vendors use such a wide variety of thermistors and connectors
for injectate temperature that Pronk Technologies cannot supply these. One solution would be to
fabricate an injectate temp simulator locally from an injectate probe and some resistors. Some
vendors use YSI 400 or YSI 700 thermistors for injectate temp. In these cases, the internal YSI
temp simulator in the SimSlim simulator can be used.