Four-ball tester

Advertisement

Caution: risk of crushing!
Never reach under the housing or the base of the four-ball tester with your hand while
installing the machine.
Afterwards align the table surface horizontally (by adjusting the feet with threaded stud) to
obtain accurate test results and ensure proper clearance of the balance scale bar.
Caution: risk of crushing!
Never reach under the table of the four-ball tester with your hand.
5.4.2 Installation of the writing utensil holder (pos. 19)
Use two hexagon bolts, M10 x 40 (pos. 87), and two discs, DIN 125, B10 (pos. 73) to
mount the writing utensil holder to the housing (pos. 1).
5.4.3 Balance scale bar (pos. 24)
5.4.3.1 Installation of the balance scale bar
Guide the balance scale bar with mounted counterweight (pos. 43) through the opening on
the left side of the housing underneath the mounted release lever (pos. 32). Ensure that the
mounted counterweight of the balance scale bar is on the same side as the release lever.
Carefully place the cutting bolt (pos. 25) into the two cutter holders (pos. 26) in the
housing.
Handle the cutters of the balance scale bars gently and with great care. Ensure that the
tips in particular are not exposed to shocks or dropped into the cutter holders to avoid
damage to the cutting geometry. Damaged cutter holders result in inaccurate test results.
Caution: risk of cutting!
The cutters at the balance scale bar have very sharp edges that can cause severe cuts.
Always wear protective gloves to install and/or transport the balance scale bar.
5.4.3.2 The balance scale bar
The VKA-110 comes with adjusted balance scale bar. The adjustment is a life-long
adjustment.
Warning! Only the manufacturer can do adjustments!