Download Print this page

HANSA Press VKA-110 Operating Instructions Manual page 13

Four-ball tester

Advertisement

2.3.2 Measuring
2.3.2.1 Measuring
The "Measurement" window allows you to enter the measurement settings.
Use "Measurement Name" to enter a new name. When you click the start button, the
suffix "@01" will be added to the name. If you do not assign any new name following the
last measurement, the name of the previous measurement will be counted up.
Warning!
If you do not assign any name, the USB storage medium will not record any
measurements.
2.3.2.2 Time-delayed recording
You can time-delay the start of the recording of a test run by clicking ON/OFF.
If the motor is running, you can start the current recording prior to the time set by clicking
the "Status" and "Recording Status" buttons. However this has no impact on the
measurement time that has already passed and the duration of the measurement.
2.3.2.3 Stop parameter: measurement
Use the "Stop Parameter: Measurement" field to set the switch-off requirements.
You can choose between two methods: switch-off after a pre-defined period of time or
switch-off after a pre-defined number of revolutions has been reached.
You can change the values (e.g. switch-off after 1,000 revolutions) by touching the
numerical values.
Mode of operation
Freely settable rotational speed
Pre-defined standard rotational speeds
Step control
13

Advertisement

loading