Part number
Various
Where used
Various
Tools and Techniques
The following tools will be required;
None
Service and maintenance check
At each service check for:
Wear and damage to assembly
Corrosion areas to tube section
Sharps
Bends and twisting of tubing
Damaged or worn, components need to be replaced, if assembled in pairs Remploy Healthcare recommends
Removal method
Depress thump trigger located on under side of mounting plate in order to release lock.
Lift and remove from frame and discard
Fitting method
Offer replacement handed swinger bracket tube sub assembly to the front tube.
Locate moulding into front tube and rotate as far as possible so that the swinger bracket tube is in line with the
side frame.
Release the thumb trigger so that the bracket locks in position.
Checking method
Twist footrest from side to side to ensure it is locked into position.
www.remployhealthcare.com
Post code S41 8NJ
Weight excluding packaging: 220grams per handed assembly (excluding footplate and stem sub assembly
R Healthcare Technical Information Manual - Dash
Fitting Instruction – footrest bracket assembly
replace both assemblies, for continuity of wheelchair in use.
Spares and Options should only be fitted by a trained and competent person
Page 48
March 2007 Revision B