Page 1
TECHNICAL INFORMATION MANUAL (Incorporating technical, service, spares and accessory information) Dash Series Wheelchairs DASH LITE 2 AP – Attendant Propelled DASH LITE 2 SP – Self Propelled April 2014 R Healthcare Technical Information Manual - Dash November 2013 Revision B...
Service Technicians and Healthcare specialists, to enable technical, service and maintenance information to be available and informative, at time of publication. Where there are omissions, errors or areas of improvement, R Healthcare would be Important pleased for the Service Technicians and Healthcare specialist’s advice in such instances.
Page 3
Push Handles 11.0 Armrests 12.0 Upholstery and trim 13.0 Wheelchair Service Record Fitting Instructions Also please refer to User guide Also please refer to Transportation guide Important R Healthcare Technical Information Manual - Dash November 2013 Revision B...
This edition of the R Healthcare Wheelchair Information Manual is an intrinsic part of our structured product information programme. We hope that the information it provides is useful and...
WARRANTY POLICY 1. It is the R Healthcare policy to guarantee our products for a period of “24” months from their placement into service. The warranty applies to any defects in workmanship or materials, which renders the product unfit for its intended use.
Page 6
1.0 Data Sheet 12.9KG 13.0KG 13.0KG R Healthcare Technical Information Manual - Dash November 2013 Revision B...
Clinical Assessment & Service Check Recommendations The Dash series is a manual wheelchair and is intended to meet the needs of a clinical assessment. The tubular folding Aluminium frame, is available in various sizes and build configurations to meet the assessed needs of the user.
Page 8
Occupant or Attendant Dash wheelchairs suitable for use with Dash spares or USER accessories, supplied by R Healthcare Ltd. Other adaptations may be suitable dependent upon clinical assessment SUITABILITY Occupant or Attendant Dash wheelchairs suitable for people who require an...
Page 9
R HEALTHCARE A controlled service programme will improve safety and performance of wheelchairs. Lack of maintenance may affect warranty. Signed............Rehabilitation Specialist. 00 08 73 00. MAR 2003. Rev B R Healthcare Technical Information Manual - Dash November 2013 Revision B...
Page 10
R Healthcare Technical Information Manual - Dash November 2013 Revision B...
When a doubt exists, where the wheelchair has suffered constant miss use of storage, the wheelchair should be removed from the client, and disposed of. R Healthcare Technical Information Manual - Dash November 2013 Revision B...
Page 12
On each Footrest hanger, disengage the latch, rotate the hanger to the rear of the chair, check that the screws on the front frame are tight using fingers only. R Healthcare Technical Information Manual - Dash Page 12 March 2007 Revision B...
Page 13
(0.030” – 0,75mm total). Any damaged castor must be replaced, R Healthcare recommend these are replaced in pairs. On each Castor in turn grip the wheel firmly through the spokes and attempt to rock the wheel on its bearings looking for excess movement of the wheel and also excess lateral movement of the wheel along its centre pivot bush.
R Healthcare encourages the use of computer data as a means of improving the effectiveness of wheelchair service providers in keeping a track of wheelchairs and the users.
Preparation for high quality repainting of wheelchair Frames, as is required in a remanufacturing operation requires the removal of all previous paint. R Healthcare have found that chemical stripping is the most successful method of paint removal, as this does not affect the structure and function of the substrate steel structure.
A component strip down and repaint may require the removal of caution and information labels. R Healthcare have reduced direct chair labelling to a minimum for practical reasons, but as a result of our experiences of actual incidents in the field, we believe that some important permanent caution labelling to particular features of a wheelchair is still necessary to remind users of important safety aspects.
Many standard service routines are applicable to all of R Healthcare wheelchair models, both Attendant and Occupant Controlled, and are detailed in this manual.
CASTOR HEAD PIVOT DASH LIFE, DASH LITE, STOWAWAY SET TO SUIT ROTATION 19mm A/F HEX SOCKET CASTORS TO FRAME MANUALLY PROPELLED BASIC FIXED LOCTITE 270 24mm A/F HEX SOCKET R Healthcare Technical Information Manual - Dash Page 18 March 2007 Revision B...
Page 19
In this application and environment the wheelchair would be occupied for long periods of time, and typically, frequency of service checks could be six months. R Healthcare Technical Information Manual - Dash Page 19 March 2007 Revision B...
Page 20
Where spares are advised as LH and RH these definitions are taken from a seated wheelchair occupant aspect. R HEALTHCARE RECOMMENDS THAT RECONDITIONING OR SERVICE OF SPARE PARTS AND ACCESSORIES IS NOT CARRIED OUT, THE ADVICE IS TO PURCHASE NEW APPROVED R HEALTHCARE PARTS.
The wheels used on Dash have been specifically developed for the intended area of application. They include several features not previously available on basic R Healthcare wheelchairs. All Dash wheelchairs have QD (Quickly detachable) rear wheels. The QD hub and spindle on Dash is 12mm diameter.
The brakes fitted to a basic Dash are parking brakes, to be applied when the chair is not in motion. R Healthcare Technical Information Manual - Dash Page 22...
Page 23
See section 3.12.3. for details. Service staff should be aware of the need for assessment rehabilitation engineer healthcare professional prior to making such changes to a chair R Healthcare Technical Information Manual - Dash Page 23 March 2007 Revision B...
To fit anti tip stabilisers it is necessary to remove the tipping lever sleeve. The frame tube under the sleeve has a fixing hole, the anti tip assembly incorporates a quick release sprung location pin R Healthcare Technical Information Manual - Dash Page 25...
Page 26
If this anti tip releasable facility is thought to be unsuitable for a particular user, it is possible to revert to the previous R Healthcare practice of securing the anti tip in position using a more permanent M6 screw fixing.
Page 27
This front socket has a sprung loaded latch pin which can be operated by a finger release catch to lock or release the armrest frame from the front socket for side transfer. R Healthcare Technical Information Manual - Dash Page 27...
Page 28
Lap straps are standard in an original specification of Dash, and supplied subsequently as spares, as indicated below. The strap is attached by means of a bracket. R Healthcare Technical Information Manual - Dash Page 28 March 2007 Revision B...
Page 29
Maximum washing temperature 30 deg. C, using water only, to remove stubborn stains a little mild detergent is useful, but must be wiped clean with warm water. Above this temperature can damage the canvas and leave residue. If desired cleanliness is not achieved R Healthcare advises the canvas is replaced.
PUSH HANDLES HANDGRIPS PAINTWORK CUSHION UPHOLSTERY ATTACHMENTS TIE DOWN POINTS OCCUPANT BELT GENERAL CONDITION CHECKERS INITIALS DATE REMPLOY HEALTHCARE (MOBILITY) 00 08 71 00. Aug 2003. Rev E R Healthcare Technical Information Manual - Dash Page 30 March 2007 Revision B...
Page 31
DETAILS OF SPARES AND ACCESSORIES ARE AVAILABLE FROM CUSTOMER SERVICE DEPARTMENT. INFORMATION IS ALSO AVAILABLE FROM THE SPARES AND DASHORY MANUAL INFORMATION R Healthcare Technical Information Manual - Dash Page 31 March 2007 Revision B...
Page 32
Spares and Options should only be fitted by a trained and competent person www.rhealthcare.co.uk Post code S41 8NJ Weight excluding packaging: 0.6kgs max. Per tyre R Healthcare Technical Information Manual - Dash Page 32 March 2007 Revision B...
Page 33
Spares and Options should only be fitted by a trained and competent person www.rhealthcare.co.uk Post code S41 8NJ Weight excluding packaging: 0.6kgs max. per tyre R Healthcare Technical Information Manual - Dash Page 33 March 2007 Revision B...
Page 34
Spares and Options should only be fitted by a trained and competent person www.rhealthcare.co.uk Post code S41 8NJ Weight excluding packaging: 0.6kgs max. per tyre R Healthcare Technical Information Manual - Dash Page 34 March 2007 Revision B...
Page 35
Spares and Options should only be fitted by a trained and competent person www.rhealthcare.co.uk Post code S41 8NJ Weight excluding packaging: 0.5 kgs max. Per wheel R Healthcare Technical Information Manual - Dash Page 35 March 2007 Revision B...
Page 36
Loose fitting hand rims Loose or flaking paint, which may cause sharp edges. Damaged or worn, components need to be replaced, if assembled in pairs R Healthcare recommends replace both assemblies, for continuity of wheelchair in use. Method 24. Using the No 3 Pozidrive Screwdriver remove the screws that lock the hand-rim on to the wheel.
Consider the use of lap straps, as they may need fitting as part of this process. Spares and Options should only be fitted by a trained and competent person www.rhealthcare.co.uk Post code S41 8NJ Weight excluding packaging: 500 grms per canvas R Healthcare Technical Information Manual - Dash Page 37 March 2007 Revision B...
Page 38
Ensure at any burrs on the canvas screws are removed. Spares and Options should only be fitted by a trained and competent person www.rhealthcare.co.uk Post code S41 8NJ Weight excluding packaging: 500 grms per canvas R Healthcare Technical Information Manual - Dash Page 38 March 2007 Revision B...
Page 39
Ensure there is no rattle and movement in the assembled back rest assembly when fitted to wheelchair. Spares and Options should only be fitted by a trained and competent person www.rhealthcare.co.uk Post code S41 8NJ Weight excluding packaging: 1.0. Kgs max. Per wheelchair R Healthcare Technical Information Manual - Dash Page 39 March 2007 Revision B...
Page 40
It is advisable for the occupant or carer to retain the lap strap to the correct position at all times, especially in outdoor conditions. Spares and Options should only be fitted by a trained and competent person www.rhealthcare.co.uk Post code S41 8NJ Weight excluding packaging: 0.35kgs max. per belt. R Healthcare Technical Information Manual - Dash Page 40 March 2007 Revision B...
Page 41
Spares and Options should only be fitted by a trained and competent person www.rhealthcare.co.uk Post code S41 8NJ Weight excluding packaging: 50 gms per hand grip R Healthcare Technical Information Manual - Dash Page 41 March 2007 Revision B...
Page 42
Castor head rock – castor pintle must not move sideways when attached to castor boss type housing. Damaged or worn, components need to be replaced, if assembled in pairs R Healthcare recommends replace both assemblies, for continuity of wheelchair in use.
Page 43
Contaminated armpads Damaged or worn, components need to be replaced, if assembled in pairs R Healthcare recommends replace both assemblies, for continuity of wheelchair in use. 35. Remove the existing armpad from the wheelchair using the necessary tool, turning anticlockwise.
Page 44
Ensure occupant or carer is aware of using the tipping levers when negotiating steps or similar obstacles. Spares and Options should only be fitted by a trained and competent person www.rhealthcare.co.uk Post code S41 8NJ Weight excluding packaging: 50 Gms max per sleeve R Healthcare Technical Information Manual - Dash Page 44 March 2007 Revision B...
Page 45
9. Make sure there is no excessive play and that all moving components can move freely. Damaged or worn, components need to be replaced, if assembled in pairs Remploy Healthcare recommends replace both assemblies, for continuity of wheelchair in use. R Healthcare Technical Information Manual - Dash Page 45 March 2007 Revision B...
Page 46
Ensure the stabiliser is set to prescription or users assessed requirements Spares and Options should only be fitted by a trained and competent person www.rhealthcare.co.uk Post code S41 8NJ Weight excluding packaging: 204 grams per stabiliser. R Healthcare Technical Information Manual - Dash Page 46 March 2007 Revision B...
Page 47
Ensure footplate height is set to occupant requirement are symmetrical through centre line of chair. Spares and Options should only be fitted by a trained and competent person www.rhealthcare.co.uk Post code S41 8NJ Weight excluding packaging: 360grams per side R Healthcare Technical Information Manual - Dash Page 47 March 2007 Revision B...
Page 48
Spares and Options should only be fitted by a trained and competent person www.remployhealthcare.com Post code S41 8NJ Weight excluding packaging: 220grams per handed assembly (excluding footplate and stem sub assembly R Healthcare Technical Information Manual - Dash Page 48 March 2007 Revision B...
Page 49
R Healthcare recommends that because of possible finger trap areas, that an attendant locates releases or adjusts the leg rest to required position as advised by assessor. Movement of Elevating Leg rests should be controlled from the outside of the chair using both hands to steady the assembly during assembly or adjustment.
Page 50
Adjustment however should only be carried out by an authorised professional who understands the clinical requirements www.rhealthcare.co.uk Post code S41 8NJ Weight excluding packaging: 2kgs per assembly. R Healthcare Technical Information Manual - Dash Page 50 March 2007 Revision B...
Page 51
Side panels for any signs of damage Armrest pads for damage, contamination Secure location into front and rear armrest pivot mouldings Damaged or worn, components need to be replaced, if assembled in pairs R Healthcare recommends replace both assemblies, for continuity of wheelchair in use. Method 39.
Page 52
Where used Various Tools and Techniques Damaged or worn, components need to be replaced, if assembled in pairs R Healthcare recommends replace both assemblies, for continuity of wheelchair in use. Fitting Turn Tray (1) upside down and secure pivot mounting bracket (2) to the tray using screw (3) washers (4,5) into pre-drilled holes (using holes that suit the chair width) Remove armrests from chair, loosen arm pad screws from armrest and remove.