R Healthcare Dash Series Technical Information Manual page 46

Incorporating technical, service, spares and accessory information, attendant propelled/self propelled
Table of Contents

Advertisement

Part number
Various
Where used
Various
Tools and Techniques
The following tools will be required;
None needed
Service and maintenance check
At each service check for:
Bent or damaged tube
Broken nylon wheels
Jammed or poor locating retaining clips – retains stabilisers to lower frame.
Sharp protrusions
Damaged or worn, components need to be replaced, if assembled in pairs Remploy Healthcare recommends
Removal method
Remove locking pin
Carefully pull away from the back of the wheelchair until free from the bottom tube, the stabilisers should now
be discarded.
Fitting method
Fit new stabilisers with the wheels pointing down locate the stabiliser into the lower tube at the rear of the
wheelchair
Insert quick release locking pin through pre-drilled hole in frame and through stabiliser tube.
Checking method
Ensure the stabilisers are locked to position.
Carefully hold the stabiliser and pull away from the wheelchair.
Ensure the stabiliser is set to prescription or users assessed requirements
www.rhealthcare.co.uk
Post code S41 8NJ
Weight excluding packaging: 204 grams per stabiliser.
R Healthcare Technical Information Manual - Dash
Fitting Instruction – Anti tip stabilisers - pair
replace both assemblies, for continuity of wheelchair in use.
Spares and Options should only be fitted by a trained and competent person
Page 46