Statistics printer
Table of Contents

Advertisement

Message/Error
W r o n g
f o r m a t !
W r o n g
m e a s . u n i t !
N o m e a s. v a l u e
a v a i l a b l e !
N o m e a s. s e r i e s
a v a i l a b l e !
n o t p o s s i b l e !
Cause
The measuring value may include a
maximum of 6 places after the decimal
point. The maximum number of digits
including the prefix and the decimal point is
10 digits. If this number is being exceeded,
an error message will be displayed.
The measuring unit of the measuring value
is selected in the set-up mode. When a
measuring instrument however is also
sending the measuring unit and this
deviates from the selcted unit, an error
message will appear.
When a measuring value is being
requested with
but is not
DATA
available within 3 sec, this error message
appears.
1.There is no measuring sequence
available for printing.
2. There is no measuring sequence
available for sending.
The required action is not possible, e.g.
attempting to change the sample size within
a started series of measuring will result in
an error message.
Remedy
Check on the measuring instrument that the
format conforms to the default.
Change the weighing unit at the YKT.
Check the connection to the interface.
Collect a new series of measuring values
e.g. abort and delete the series of
measurements. Subsequently alter the
sample size.

Hide quick links: