Technical data

17 Technical data

17.1 Technical data for circulator

Circulator
Working temperature range
Temperature stability
Temperature setting
Temperature display
Resolution
ATC - Absolute Temperature Calibration
Temperature control
Heating capacity (at 100 V / 50 Hz / 60 Hz)
Heating capacity (at 115 V / 50 Hz / 60 Hz)
Heating capacity (at 200 V / 50 Hz / 60 Hz)
Heating capacity (at 230 V / 50 Hz / 60 Hz)
Circulating pump:
Delivery rate at 0 bar
Pressure at 0 liters
Max. viscosity
Dimensions (WxDxH)
Useful immersion depth
Weight
Ambient temperature range
Mains power connection 100 V/50 Hz/60 Hz
Power consumption
Mains power connection 115 V/50 Hz/60 Hz
Power consumption
Mains power connection 200 V/50 Hz/60 Hz
Power consumption
Mains power connection 230 V/50 Hz/60 Hz
Power consumption (at 230 V)
Classification, according to DIN 12876-1
°C
°C
°C
kW
kW
kW
kW
l/min
bar
cSt
13.2 x 16.0 x 35.5
cm
cm
kg
°C
V / Hz
100 ±10 % / 50 / 60
A
V / Hz
115 ±10 % / 50 / 60
A
V / Hz
200 ±10 % / 50 / 60
A
V / Hz
230 ±10 % / 50 / 60
A
DYNEO
DD
20 ... 200
±0.01
digital
TFT
0.01
3-point
PID3
0.8
1.0
1.5
2.0
8...23
0.1...0.6
50
16.0
2.5
5 ... 40
10
11
9
10
III (FL)
73