Emerson Rosemount 327T Quick Start Manual page 22

Temperature transmitter
Table of Contents

Advertisement

Troubleshooting
January 2019
Fnr
91
92
100
22
Type of fault
Supply voltage is outside the operating
voltage range.
Operating temperature of the electronics is
outside the recommended range.
Error while setting parameters using IO-
Link.
Quick Start Guide
00825-0100-4329
Corrective measures
Check the supply voltage and ensure the
voltage supply is correct.
1. Check the temperature of the
transmitter housing considering
ambient and process temperature
conditions.
2. Make necessary adjustments to
adhere to the recommended
operating temperature range.
Modify the parameter setting using IO-Link
again or reset all parameters to factory
default settings.
Quick Start Guide

Advertisement

Table of Contents
loading