8 Fault Analysis/Trouble-shooting
8.2.11 Analogue outputs configuration error
8.2.12 Service warning 1 / service warning 2
Analogue outputs are parameterised that cannot physically be found. Check
parameters 539, 549, 559, 569 and 530, 540, 550,560 and compare with the inserted
cards. If necessary, replace analogue output cards and/or the processor board.
The service warning draws attention to the need for regular servicing. The user can
define service warnings as required, e.g.
Service warning 1 → check probe
Service warning 2 → replace probe
The appropriate cycle times can be configured as desired by means of parameters
1260 and 1261.