Table of Contents

Advertisement

Quick Reference for End Users
Lambda Transmitter LT2-K
Lambda Probe LS2-K/LS2KV
Sensors and Systems for Combustion Engineering
www.lamtec.de

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Lambda LS2-K and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Lamtec Lambda LS2-K

  • Page 1 Quick Reference for End Users Lambda Transmitter LT2-K Lambda Probe LS2-K/LS2KV Sensors and Systems for Combustion Engineering www.lamtec.de...
  • Page 3: Table Of Contents

    Table of Contents Table of Contents Important Information about the Manual ........5 Validity of these Instructions.
  • Page 4 Table of Contents 4.5.4 LT2-K Parameter Menu ......... . . 39 Factory Settings .
  • Page 5 Table of Contents 8.5.6 Value Is Incorrect ..........68 8.5.7 Reference Air Is Missing .
  • Page 6: Important Information About The Manual

    Important Information about the Manual Important Information about the Manual Validity of these Instructions This manual describe the Lambda Transmitter LT2 with all required components. The informa- tion in this document applies to the software version 1V61. If you utilise a different version, this can lead to other effects to your device than those described in this manual.
  • Page 7: For Your Safety

    For Your Safety For Your Safety Classification of the Safety Instructions and Warnings The following symbols are used in this document to draw the user's attention to important safe- ty information. They are located at points where the information is required. It is essential that the safety information is observed and followed, and that applies particularly to the warnings.
  • Page 8: Proper Use - Conditions Of Use

    For Your Safety Proper Use - Conditions of Use The Lambda Transmitter LT2 in combination with the Lambda Probe LS2-K, is a universal sys- tem for a direct detection of combustible gas constituents (CO/H ), displayed as an equivalent of CO (CO ) in combustion systems in superstoichiometric range (...
  • Page 9: Safety Equipment/Safety Measures

    User groups For the handling of the Lambda Transmitter LT2, three user groups are required: • Service technicians of LAMTEC or its OEM customers and/or trained customer personnel: Qualified technicians/engineers  have very good knowledge of the device. – –...
  • Page 10: Protection Against Emissions From Gas Carrying Channels

    For Your Safety Protection Against Emissions from Gas Carrying Channels Protect against gas leakage from the gas-bearing channel The Lambda Probe LS2-K is fitted directly to the gas-bearing channel via a counterflange. If the Lambda Probe LS2-K is dismantled, depending on the plant but particularly in the case of excess pressure, aggressive and / or hot gas can seep out of the channel and this could pose a serious health hazard to an unprotected operator if suitable safety measures have not be taken previously.
  • Page 11: Commissioning/Decommissioning

    For Your Safety Commissioning/Decommissioning 2.6.1 Decommissioning, Storage and Commissioning The Lambda Transmitter and the Lambda Probe form a high quality electronic system. Treat them with care at all times, including during shut-down, transport and storage. Shut-down NOTICE Do not switch the Lambda Transmitter off as long as the Lambda Probe is mounted; including when the relevant facility has been shut down.
  • Page 12: Technical Description

    Technical Description Technical Description LT2-K LS2-K Structure in Principle Fig. 3-1 LT2 LS2-K system components. Pre-filter type 650R2055 Counter flange DN65PN6 type 655R0179/0180 Lambda Probe LS2-K with semi-automatic calibration unit type 656R1030/1031/1032 Probe connection box (PCB) Connection for offset/test gas calibration – 0.3 bar instrument air, test gas for example: 3 vol.
  • Page 13: Lt2-K Ls2-Kv Structure In Principle

    Technical Description LT2-K LS2-KV Structure in Principle Fig. 3-2 System overview LT2-K LS2-KV Complete probe with counter flange and PCB LT2-KV Lambda Transmitter in wall mounting housing IP65 with integrated reference air pump (option), type 657R102, steel sheet LS2-KV Lambda Probe with fully automatic calibration unit type 650R2050/2051/650R2052 Sintered metal pre-filter type 650R2055 Counter flange DN65PN6 type 655R0179/0180 Flange seal Klinger Sil C-4400 type 655P4209...
  • Page 14: Ls2-K / Ls2-Kv Lambda Probe

    Technical Description LS2-K / LS2-KV Lambda Probe The wiring between the LS2-K / KV Lambda Probe and the LT2-K Lambda Transmitter is done with conventional, screened cables over terminal blocks; see the wiring diagram in chapter 10.2 Connection Diagram for LT2-K/LS2-K (semi automatic calibration)10.3 Connection Dia- gram for LT2-KV/LS2-KV (fully automatic calibration) 1 Probe connection box (PCB) Fig.
  • Page 15 Technical Description 1 Probe connection box (PCB) Fig. 3-5 Lambda Probe LS2-KV with connection for fully-automatic calibration 1 Absolute pressure sensor 2 Differential pressure sensor 3 Solenoid valve 2 (shut-off valve) 4 Throttle for calibration gas 5 Solenoid valve 1 (shuttle valve) 6 Throttle for reference air Hose connection for reference air (instrument air 0.3 bar)
  • Page 16: Lt2-K Lambda Transmitter

    PC and for connecting to other LAMTEC devices. A universal BUS interface for connecting to customers control systems is available as an option.
  • Page 17: Lt2-K Lambda Transmitter Type 657R102 With Display And Operating Unit

    Transformer probe and elec- tronic Connection for Remote-Dis- play-Software (Option) Connecting cable with plug for display and operating unit Electrical connections BUS-interface alternative LAMTEC SYSTEM BUS (CAN-BUS) Operating mode display multifunction push button maintenance switch Monitor output Protective earth terminal for cable screening...
  • Page 18: Display And Operating Unit

    Technical Description 1 - L 2 - N 4 - PE Fig. 3-11 Connector Fig. 3-10 Connector for power supply NOTICE The dimensions of the wall mounting housing changes with the selection of the option ’with reference air pump’. Display and Operating Unit 1 = Cursor keys 2 = ENTER key 3 = Menu keys...
  • Page 19: Cold-Start Delay

    Technical Description Cold-start Delay Coldstart delay is used for suppressing incorrect measured values while the probe is heating up. A cold start delay is activated always after ’Power off’ or the replacement of a probe. The cold start delay can be aborted at any time: •...
  • Page 20: Digital Outputs

    NOTICE To avoid EMC faults shield the branch lines of the relay modules. When using unshielded cables loop them twice trough the provided ferrite ring. Options for LAMTEC Combination Probes – Semi-automatic calibration (available for probes with attribute ’K’ ) –...
  • Page 21: Analogue Inputs (Optional)

    Technical Description 3.10 Analogue Inputs (Optional) Parameter groups 570 ... 609 By plug-in card on the electronic unit of the LT2-K power supply (max. 2) • Analogue input card 0/4 ... 20 mA type 663P6001 • Analogue input card 0/4 ... 20 mA with 24 VDC supply for transmitter for LT1/LT2 type 663P6002 •...
  • Page 22: Commissioning/Decommissioning

    Commissioning/Decommissioning Commissioning/Decommissioning Display and Operating Unit of the LT2-K Lambda Transmitter The LT2's operation and the display of measured values, operational and error messages take place by the (optional) display and operating unit, or by a PC in combination with the Remote Display Software.
  • Page 23: Internal Display And Operating Elements

    Commissioning/Decommissioning Internal Display and Operating Elements Legend: LED is off flashes lights up Multifunction key T 2 Operation (green) LED 6 Operating mode (green) LED 5 Maintenance switch S 1 Operation Measurement Calibration Offset compensation (flashes slowly) With test gas/comparative measure- ment (flashes quickly) Maintenance mode off Maintenance mode on...
  • Page 24: Display And Operating Unit Type 657R0831 (Option)

    Commissioning/Decommissioning Display and Operating Unit Type 657R0831 (Option) Fig. 4-2 Display of LT2-K with key assignment shows the display of the LT2 Lambda Transmit- ter. It consists of a LCD display, the cursor keys, the ENTER key, and the menu keys. NOTICE - The cursor keys are used to select measurement values, parameters or functions in the visual area of the display.
  • Page 25: Display

    Commissioning/Decommissioning 4.4.1 Display The display is devided into four areas: (see Fig. 4-3 Arrangement of the display): 1. The status line at the top of the display shows the following: • On the left side, whether the maintenance mode is activated or not •...
  • Page 26: Menu Function 'Meas

    Commissioning/Decommissioning 4.4.2 Menu Function 'meas' When pressing the [ meas ] the display switches to a large representation of the measure- ment value that has been selected with the cursor keys (up, down) (see 4.4 Display and Op- erating Unit Type 657R0831 (Option)). Pressing the [ meas ] key repeatedly switches the display back to a listing of all the measured values.
  • Page 27 Commissioning/Decommissioning During manual calibration the calibrated value can be change by the menu function ’Change of calibration value probe LS2-K/LS2-KV’. Fig. 4-5 Calibration functions Fig. 4-6 Menu functions After calling up this menu function, the amendment procedure in the displayed sub-menu can be terminated with the [ cancel ] menu key.
  • Page 28: Menu Function 'Par

    Commissioning/Decommissioning 4.4.4 Menu Function 'par' After calling up the [ par ] menu the menu to set parameters opens. Fig. 4-8 Password entry Fig. 4-7 Parameter menu Access to the parameter menu is divided into the following authorization levels: • Operation level Customer level ...
  • Page 29: Menu Function 'Psw

    Commissioning/Decommissioning 4.4.5 Menu Function 'psw' [ psw ] calls up the password input menu. The authorization level* is displayed. The functions which a in the menu are described below: • [exit] returns to the [ par ] menu. • [ clear ] resets the access level to ’Operation’ *. •...
  • Page 30 Commissioning/Decommissioning Display long: – The same as Medium, but with an addi- tional parameter status line. [ group- ] changes one parameter group back. [ group+ ] changes one parameter group forward. All available parameter groups are listed in the appendix. The left/right cursor keys correspond to the key-function of [ group - ] and [ group + ].
  • Page 31: Menu Function 'Change

    Commissioning/Decommissioning 4.4.7 Menu Function 'change' [ change ] allows parameter values to be amended. The menu bar for the sub-menu is the same as the [ view ] menu (see chapter 4.4.6 Menu Function 'view'). Unlike this the parameter which should be changed is displayed inversely (bright letters on dark ground).
  • Page 32: Menu Function 'Diag

    Commissioning/Decommissioning 4.4.8 Menu Function 'diag' On pressing the [diag] key, the display switches to warnings and faults. Use the up/down cur- sor keys to select individual warnings or faults, or limit values. NOTICE Limit values are displayed only if they are also activated by parameters 930/940/950/960 (ser- vice level).
  • Page 33: Limit Values Display And Resetting

    Commissioning/Decommissioning Description: Limit value 1 is set by parameter to O measurement value. Switching point: exceeding 10.0% per volume O short fall 3.0 % per volume O Recent O measurement value 6.7 % per volume O The limit value is not set. Limit value2 has responded.
  • Page 34: Display Of The Limit Value Crossing

    Commissioning/Decommissioning The condition ’inactive’ is displayed, if the limit value is not set in the parameter or if the limit value is deactivated by the digital inputs or the operating mode (parameter 967). • All limit values can be deactivated together depending on the operating mode. The behaviour can be set in parameter 967 in service level.
  • Page 35: Brightness, Contrast And Language

    Commissioning/Decommissioning 4.4.11 Brightness, Contrast and Language The parameter group ***Display*** controls the LT2 Lambda Transmitter's display and oper- ating unit with the following parameters: • P 970 Contrast Adjust the display and operating unit’s contrast. • P 971 Brightness Adjust the display and operating unit’s background brightness •...
  • Page 36: Customer Password Input

    NOTICE The customer gets the password for service access level by a letter. Only trained personnel may get this password. OEM customers have their own company-specific password. This is the same in all LAMTEC devices.
  • Page 37: Menu Overview

    Commissioning/Decommissioning Menu Overview 4.5.1 LT2-K Calibration Menu Fig. 4-13 Overview of the calibration menu cal...
  • Page 38: Lt2-K Diagnosis Menu

    Commissioning/Decommissioning 4.5.2 LT2-K Diagnosis Menu Fig. 4-14 Overview diagnosis (diag)
  • Page 39: Lt2-K Measurement Menu

    Commissioning/Decommissioning 4.5.3 LT2-K Measurement Menu Fig. 4-15 Overview measurement menu (meas)
  • Page 40: Lt2-K Parameter Menu

    Commissioning/Decommissioning 4.5.4 LT2-K Parameter Menu Fig. 4-16 Overview parameter menu (par)
  • Page 41: Factory Settings

    Commissioning/Decommissioning Factory Settings 4.6.1 Device Configuration (unless otherwise specified in the order) Measurement range: 0. .. 30 % volume of O Resolution: 0.1 % volume of O in the range of 0 ... 18 % volume of O 1.0 % volume of O in the range higher than 18 % volume of O Probe temperature: 1000 K (parameter 141)
  • Page 42: Measurement Start-Up

    Commissioning/Decommissioning Measurement Start-up NOTICE When installing the probe and during subsequent operation, it is necessary to make sure that the probe does not get into contact with oil, grease or with agents used to clean the tank. This belongs to the cell and the connection area! Poisoning or contamination of the probes is indicated by a potential difference across the air gap of <...
  • Page 43: Measurement With Ls2-K

    Commissioning/Decommissioning 4.7.1 Measurement with LS2-K 1. Connect the probe electrically and pneumatically, see connecting diagram chapter 10.1 Electrical Connectionand 10.4 Electrical Connection. 2. Supply reference air connection and calibration gas connection with instrument air of 0.3 bar. NOTICE If instrument air is not available use regular compressed air for air calibration. If neither instru- ment air nor compressed air is available proceed one of the following options: •...
  • Page 44 Commissioning/Decommissioning NOTICE Function Key operation Toggle the displayed warning/ fault Press briefly Reset the displayed warning/fault Press for longer than 3 s* Abort cold-start Press for longer than 3 s** Trigger an offset calibration to ambient air Press key for longer than 3 during measurement** * Some warnings and faults cannot be reset if the error is still present or the routine is still running.
  • Page 45: 4.7.1.1 Adjust The Amount Of The Reference Air

    Commissioning/Decommissioning 4.7.1.1 Adjust the Amount of the Reference Air NOTICE Only oil-free and water-free compressed air can be used as reference air. Recommendation: Use instrument air if available. If neither instrument air nor compressed air is available a LT2 with integrated reference air pump is necessary.
  • Page 46: 4.7.1.2 Performing An Offset Calibration To Ambient Air

    Commissioning/Decommissioning Differential pressure sensor Absolute pressure sensor Throttle for calibration gas Throttle for reference air Stopcock for calibration gas Hose connection for test gas (instrument air or test gas ca. 0.3 bar) Outlet for air Hose connection for reference air (instrument air ca.
  • Page 47: 4.7.1.3 Performing An Offset Calibration With Compressed Air

    Commissioning/Decommissioning NOTICE Function Key operation Toggle the displayed warning/ fault Press briefly Reset the displayed warning/fault Press for longer than 3 sec* Abort cold-start Press for longer than 3 sec** Trigger an offset calibration to ambient air Press key for longer than 3 sec during measurement** * Some warnings and faults cannot be reset if the error is still present or the routine is still running.
  • Page 48: 4.7.1.4 Perform A Calibration With Reference Measurement

    Commissioning/Decommissioning 4.7.1.4 Perform a Calibration with Reference Measurement Select ’Calibration/test with reference-measuring’ in the [cal] menu and confirm with ENTER. Use the up/down cursor keys to adjust the O value of the reference measurement. Confirm with [OK] key (see Fig. 4-21 Input of the calibration values). This is valid only for O values smaller than 18 %.
  • Page 49: 4.7.1.5 Perform A Calibration With Test Gas

    Commissioning/Decommissioning 4.7.1.5 Perform a Calibration with Test Gas NOTICE The calibration can be proceeded with test gas as an alternative to the reference measure- ment. The procedure corresponds to the offset calibration with compressed air. Fig. 4-22 Display of calibration with test gas NOTICE The test gas used must be free of oxidizing components such as CO, NO, etc.Oxidising com- ponents are burned at the 700 °C hot Pt electrode catalytically and distort the calibration.
  • Page 50: Ls2-Kv Measurement (Fully-Automatic Calibration)

    Commissioning/Decommissioning 4.7.2 LS2-KV Measurement (Fully-automatic Calibration) • Apply instrument air of 0,3 bar at the probe’s reference air connection and test gas 1 con- nection. Apply test gas of 2 … 5 % volume of O in N of 0,3 bar on test gas 2 connection. CAUTION! The test gas used must be free of oxidizing components such as CO, NO, etc.
  • Page 51: 4.7.2.1 Adjust The Amount Of The Reference Air

    Commissioning/Decommissioning 4.7.2.1 Adjust the Amount of the Reference Air Differential pressure sensor Absolute pressure sensor Solenoid valve 2 (shut off valve) Throttle for calibration gas Solenoid valve 1 (shutter valve) Throttle for reference air Hose connection for reference air (instrument air 0.3 bar) Outlet reference air Hose connection test gas 1 (instrument air 0.3 bar)
  • Page 52: 4.7.2.2 Performing An Offset Calibration To Ambient Air

    Commissioning/Decommissioning 4.7.2.2 Performing an offset calibration to ambient air Press the [ cal ] menu key to display a choice two calibration variants Fig. 4-24 Selection menu of offset calibration NOTICE To adjust the required rise of the pressure during calibration manual calibration must be se- lected.
  • Page 53: 4.7.2.3 Reference Measurement - Performing A Test Calibration

    Commissioning/Decommissioning 4.7.2.3 Reference measurement - Performing a Test Calibration Perform the calibration by: • Display and operating unit • Remote Display Software Press the [cal] menu key to display a choice two calibration variants (see chapter 4.7.2.2 Per- forming an offset calibration to ambient air). NOTICE Selecting the manual calibration in CALIBRATION mode runs a calibration with reference measurement and test gas as described in chapter 4.7.1.4 Perform a Calibration with Refer-...
  • Page 54: 4.7.2.4 Performing A Combined Offset And Test Gas Calibration

    Commissioning/Decommissioning 4.7.2.4 Performing a Combined Offset and Test Gas Calibration Start calibration alternatively by: • Display and operating unit • Remote Display Software • Digital input 5 • Cyclic (parameter 300) Pressing the [cal] menu key shows a choice of 2 calibration variants. (see figure in chapter 4.7.1.2 Performing an offset calibration to ambient air).
  • Page 55: Probe Certificate

    Commissioning/Decommissioning Probe Certificate There is a probe certificate included in every probe. This should be completed as part of the commissioning work. The probe certificate must always be included in the event of any com- plaints or repairs. NOTICE The settlement of problems on the basis of goodwill requires the submission of the passport. Fig.
  • Page 56: Setting Up Service Warnings

    Commissioning/Decommissioning Setting up Service Warnings The service warnings are designed to draw attention to regular servicing. The service warn- ings can be freely defined by the operator, e.g. Service warning 1  check probe Service warning 2  replace probe The appropriate cycle times can be freely configured by the parameters 1260 and 1261.
  • Page 57: Test Certificate

    Test Certificate Test Certificate Each probe has an accompanying production test record. This means current measurement values can be compared with the bench performance at any time. Fig. 5-1 Example test protocol...
  • Page 58: Operation

    Operation Operation Measurement During Distinct Pressure Surges at the Measuring Site If the display jumps badly, damping can be increased via the (optional) display and operating unit and the service and diagnostic software (i.e. by increasing the integration's time-con- stant); this steadies the display: parameter 360 - Operational release level. However, this slows down the display in terms of reaching an end state.
  • Page 59: Service And Maintenance

    Service and Maintenance Service and Maintenance Checking the O Probe 7.1.1 Calibrating the Air Voltage (Offset) Requirements: Measuring cell in air or applied with air. • Connect compressed air/instrument air to the PCB of the LS2-K • The procedure for LS2-K see chapter 4.7.1.2 Performing an offset calibration to ambient air / Fig.
  • Page 60: Fully-Automatic Combined Calibration

    Service and Maintenance 7.1.3 Fully-automatic Combined Calibration • The procedure for LS2-KV see chapter 4.7.2.4 Performing a Combined Offset and Test Gas Calibration. Checking the Measurement Input of LT2-K Connect a digital voltmeter between terminals 33 (-) and 34 (+), in parallel to the probe. Com- pare the measured voltage with the probe voltage displayed (U Range: -30 mV …...
  • Page 61: Cleaning/Replacing The Sintered Metal Pre-Filter

    Service and Maintenance 7.4.3 Cleaning/Replacing the Sintered Metal Pre-filter Cleaning the filter is only necessary if its permeability has been impaired due to deposits. In this case the following warnings appear: – ’Delta-P offset calibration too high, filter clogged’ – ’Delta-P offset calibration too low, too little gas’...
  • Page 62: Replacing The O Measuring Cell

    Service and Maintenance 7.4.4 Replacing the O Measuring Cell To replace the O measurement cell the repair kit 500 mm 650R2060 for measurement cells is necessary: 1000 mm 650R2061 1500 mm 650R2062 • disconnect LS2-K/LS2-KV Lambda probe from electrical and gas supplies and remove the probe •...
  • Page 63 Service and Maintenance Fig. 7-3 Drawing probe Quantity Description Material Allen key 1.4571 Filter SIKA H 20 Metal O-ring  31,8 mm (1.252" in) Inconell 600 Metal O-ring  46,2 mm (1.819" in) Inconell 600 sensor incl. mount and rod ceramics Hex nut 1.4571...
  • Page 64: Fault Analysis/Trouble-Shooting

    (Option) in the [diag] menu Remote Display Software LAMTEC SYSTEM BUS on external operating unit. LED row, LED 7 to12on the processor board in the LT2. If several faults, warnings are present, they can be called in sequence by activating the T2 multifunction key.
  • Page 65: Warning Indication By Led Line

    Fault Analysis/Trouble-shooting Warning Indication by LED Line Indication by LED rows, LED 7 to 12, LED 12 ON = warning LED 12 Warning No.: no warning Internal resistor probe 1 too high Fig. 8-2 Indication of the Offset outside the limits active warnings (red) LED is ON permanently Reference air is missing...
  • Page 66: Resetting Faults/Warnings

    Fault Analysis/Trouble-shooting LED 12 Warning No.: Service warning 2 Probe dynamic is missing Dynamic test is triggered Only relevant in combination with integrated O control. control the dynamic test has to be switched off parameter 1330 0) Without integrated O Resetting Faults/Warnings •...
  • Page 67: Faults - Causes And Solutions

    Fault Analysis/Trouble-shooting Faults - Causes and Solutions 8.5.1 Probe voltage too low Probe + / - reversed probe connection terminal 33-34 change • Probe poisoned replace • NOTICE Probe voltage to air +5 to –30 mV 8.5.2 Probe Heating Defective NOTICE In 99 % of all cases, the LS2-K Lambda Probe was disconnected during operation.
  • Page 68: No Probe Dynamics

    Fault Analysis/Trouble-shooting 8.5.4 No Probe Dynamics No probe dynamic detected. Check the probe. NOTICE Testing is shut off on delivery. Activate testing in parameter group 1330 to 1334. The testing checks, if the measuring value changes more than the parameter set threshold value during a specified time.
  • Page 69: O 2 Value Is Incorrect

    Fault Analysis/Trouble-shooting 8.5.6 Value Is Incorrect If a control measurement results in a different O value than the one displayed: • Has the relationship between wet/dry measurement been taken into account? see chapter 10.8 Wet/Dry Measurement Deviations, Conversion Table • Check LT2 and probe, see chapter 7.2 Checking the Measurement Input of LT2-K •...
  • Page 70: Warnings - Causes And Solutions

    Fault Analysis/Trouble-shooting Warnings - Causes and Solutions In general: Warnings do not affect the measurement functions. 8.6.1 LT2-K Calibration Menu This message is output if the AC internal resistance (R ) of the ZrO cell exceeds the permis- sible limit of 200 during operation. Possible cause: Probe aged (worn out) ...
  • Page 71: Delta-P Offset/Test Gas Calibration Is Insufficient - Too Less Gas

    Fault Analysis/Trouble-shooting 8.6.4 Delta-P Offset/Test Gas Calibration is Insufficient - Too less Gas Insufficient rising of the pressure during offset calibration or test gas calibration at LS2-KV. Possible causes: • Instrument air or test gas (1 bar) is missing at the probe. •...
  • Page 72: Test Gas Calibration Is Invalid

    Fault Analysis/Trouble-shooting 8.6.7 Test Gas Calibration Is Invalid The probe temperature which is calculated from the probe voltage and the concentration of gas is outside a range of 800° - 1200° Kelvin. Possible causes: • The connected test gas has not the same O concentration as set in parameter 301.
  • Page 73: Spare Parts

    Spare Parts Spare Parts All relevant spare parts are listed below. Recommendation: Keep the parts marked with in stock. Keeping parts marked with in stock is discretionary. Keeping parts marked with in stock only makes sense if system is equipped with the corre- sponding option.
  • Page 74: Appendix

    Appendix Appendix 10.1 Electrical Connection Fig. 10-1 Connection diagram LT2 Only in type 657R103 via terminals in type 657R102 and 657R104 via plug. The max. current burden for all 4 measurement cards is together: 80 mA...
  • Page 75: Connection Diagram For Lt2-K/Ls2-K (Semi Automatic Calibration)

    Appendix 10.2 Connection Diagram for LT2-K/LS2-K (semi automatic calibration) Fig. 10-2 Connection diagram for Gerät/Variable1 semi automatic calibration...
  • Page 76: Connection Diagram For Lt2-Kv/Ls2-Kv (Fully Automatic Calibration)

    Appendix 10.3 Connection Diagram for LT2-KV/LS2-KV (fully automatic calibration) Fig. 10-3 Fully automatic calibration LT2/LS2-KV...
  • Page 77: Electrical Connection

    Appendix 10.4 Electrical Connection Fig. 10-4 Connecting variants LT2 Cable cross section: Measuring cell heating other connecting cables terminals 35/36 up to 20 m (65.61 ft)= 1.5 mm (AWG 15) 0.75 mm (AWG 19) up to 50 m (164.04 ft)= 2.5 mm (AWG 13)
  • Page 78: Jumpers

    Plug for Remote Display Software in combination with RS232 interface module DIP switch Multifunction key BR101 termination resistor for LAMTEC SYSTEM BUS BR 106 and 107 potential allocation for digital outputs Terminal for LAMTEC SYSTEM BUS Terminal for digital inputs...
  • Page 79 LAMTEC SYSTEM BUS BR101: Selection of connection resistance: 1-2 off, 2-3 on. BR105: Position 2-3 The new processor card must also be set up for the LAMTEC SYSTEM BUS (BR12 and BR13 in position “C“). Digital inputs BR106, 107: Position 1-2: Digital inputs referenced to instrument potential.
  • Page 80: Fuses

    1.25 A inertial (IEC60127-2/3 or /5) 4 A inertial Probe heating and 24 V supply from 11/2016 3,15A ultra slow fuse 0.315 A inertial (IEC60127-2/3 or /5) LAMTEC SYSTEM BUS NOTICE Fuse replacement F1/F3 when changeover 230/115 VAC necessary !
  • Page 81: Lt2 Power Electronic Type 657E1882

    Power supply connection 230/115 V, 50/60 Hz Row of LED’s for operating and status messages Module 1-4 (fault/warning) Analogue inputs (voltage, current, potentiometer) LAMTEC SYSTEM BUS Power supply switch Digital inputs (open collector) Electrical connection X211 for relay module CAUTION! Different fuse figures F1 for 230 V and 115 V supply voltage.
  • Page 82: Wet/Dry Measurement Deviations, Conversion Table

    Appendix 10.8 Wet/Dry Measurement Deviations, Conversion Table NOTICE The LT2 carries out measurements directly in the humid flue gases (wet measurement). When extractive devices are used, flue gases are removed and prepared. "Dry Measurements" are normally used here, since the humidity has been extracted from the flue gas. As a result, O measurement values vary (see diagrams below).
  • Page 83: Eu Declaration Of Conformity

    EU Declaration of Conformity EU Declaration of Conformity...
  • Page 84 EU Declaration of Conformity...
  • Page 85 EU Declaration of Conformity...
  • Page 86 The information in this publication is subject to technical changes. LAMTEC Meß- und Regeltechnik für Feuerungen GmbH & Co. KG Wiesenstraße 6 D-69190 Walldorf Telefon: +49 (0) 6227 6052-0 info@lamtec.de Telefax: +49 (0) 6227 6052-57 www.lamtec.de Printed in Germany | Copyright  2019...

This manual is also suitable for:

Lambda lt2-kLambda ls2kv

Table of Contents