Functioning control
Mag2000 PLUS equipment offers a magnet in order to control the device
functioning.
Control procedure:
1. switch on the device in according to user manual safety prescriptions;
2. start a treatment in according to user manual instructions;
3. get the magnet and put it near to applicator;
4. check magnet vibration (it will be proportional to selected treatment
frequency).
Please contact the manufacturer in case of magnet vibration absence.
Cleaning
Clean the equipment from the dust using a dry soft cloth.
Resistant stains can be removed using a sponge soaked in solution of
water and alcohol (20%).
We recommend to disconnect the applicator from the device before
cleaning the elastic therapeutic belt with 3 solenoids or the circular cases
of professional solenoids couple.
•
Extract the cable with 3 solenoids by removing the 2 silver studs
through a screwdriver or open the circular cases through lateral
zip.
•
Clean the tissue using water and mild soap and wait for the
complete drying before reconnecting the applicators.
ATTENTION: always respect the applicators polarity paying attention to
insert the bobbins with the side indicated by + symbol turned to the green
side of elastic belt (therapeutic side).
I.A.C.E.R. Srl
Maintenance
24
MNPG53-02