MSA Ultima OPIR-5 Alignment Manual page 2

Hide thumbs Also See for Ultima OPIR-5:

Advertisement

You will now begin to align the system:
1. The front fine adjustment screw tilts the unit down. Starting with the Source, lightly tighten the front fine adjustment screw until it meets
some resistance. Slowly make a small turn to tighten the screw and lower the front of the unit. Wait a few seconds for the AJ number
to settle at the Receiver. Continue making small turns, waiting after each turn for the AJ number to settle. Once the AJ number begins
to decline, unscrew the front fine adjustment screw until it is flush with the assembly surface. Always be sure to back out screw
to avoid stripping.
2. Repeat the same procedure for the back fine adjustment screw on the unit. The back fine adjustment screw tilts the unit up. Lightly
tighten the back fine adjust screw until it meets some resistance, and then slowly make a small turn to further tighten it. Wait a few
seconds for the AJ number to settle on the Receiver. Continue until the AJ reaches its maximum and begins to decline. In most cases,
the AJ won't be in the optimal range listed in the table since the Receiver hasn't been adjusted yet. Unscrew the back fine adjustment
screw until it is flush with the assembly surface. Always be sure to back out screw to avoid stripping.
3. Secure the two front cap screws. Verify that the AJ number does not decrease by more than 2 (an increase of any amount is acceptable).
If it does decrease by more than 2, you will need to repeat steps 1 and 2.
4. Adjust the Source pan nut for maximum AJ by first loosening the pan nut, moving the unit left or right to optimize the AJ number.
Slowly tighten the nut at the bottom of the pan-tilt assembly while holding the left side of the Source. Verify that the AJ does not
decrease by more than 2 (an increase of any amount is acceptable). If it does decrease by more than 2, you will need to repeat steps 1 - 3.
• With the Source now secure, repeat steps 1-4 for the tilt and pan with the Receiver.
• Once the Receiver is secure, verify that the AJ is comparable to typical values listed
in the above table.
• At the Receiver, put the magnet on the MSA logo to exit alignment mode. The unit will
display a flashing "AC" and begin zeroing the gas reading. Once zeroing is complete,
the unit will display "0C", and then return to normal operation (0). System alignment
is now complete.
• If you're having difficulty aligning the unit within the AJ values listed in the table,
you can access the Flash Input Diagnostics mode via the menu, take note of the three
values supplied and call Tech Support for further assistance.
Note: This Bulletin contains only a general
description of the products shown. While uses
and performance capabilities are described,
under no circumstances shall the products be
used by untrained or unqualified individuals
and not until the product instructions
including any warnings or cautions provided
have been thoroughly read and understood.
Only they contain the complete
and detailed information
concerning proper use and
care of these products.
ID 0782-09-MC / April 2016
© MSA 2016 Printed in U.S.A.
Source: Tilt, Pan | Receiver: Tilt, Pan
MSA – The Safety Company
1000 Cranberry Woods Drive
Cranberry Township, PA 16066 USA
Phone
724-776-8600
www.MSAsafety.com
U.S. Customer Service Center
Phone
1-800-MSA-INST
Fax
1-800-967-0398
MSA Canada
Phone
1-800-672-2222
Fax
1-800-967-0398
MSA Mexico
Phone
01 800 672 7222
Fax
52 - 44 2227 3943
MSA International
Phone
724-776-8626
Toll Free
1-800-672-7777
Fax
724-741-1553
Email
msa.international@msasafety.com
www.MSAsafety.com/detection

Advertisement

loading