Calibration Adjustment - Fluke 772 Calibration Manual

Milliamp process clamp meter
Hide thumbs Also See for 772:
Table of Contents

Advertisement

772/773
Calibration Manual

Calibration Adjustment

The Meter features closed-case calibration adjustment with a known reference
source. The Meter measures the applied reference source, calculates correction
factors, and keeps the correction factors in nonvolatile memory.
Before you start calibration adjustment, let the Meter stabilize to room
temperature.
To turn on Calibration mode:
1. Remove the batteries and substitute with a lab supply set to 6 V dc.
2. Open the battery door. The calibration button is usually hidden by the factory
3. Use a small probe and push the calibration button longer than 2 seconds.
There are five Meter functions to adjust:
1. Clamp measure
2. mA measure
3. Volt measure (773 only)
4. mA source
5. Volt source (773 only)
14
calibration seal.
See Figure 4.
Figure 4. Accessing the Calibration Button
Calibration Button
gep01.emf