Levelflex FMP55
Power supply failure
Potential equalization
Terminals
Cable entries
Cable specification
Endress+Hauser
FOUNDATION Fieldbus
Device basic current
15 mA
Failure current FDE (Fault
0 mA
Disconnection Electronic)
FISCO
U
17.5 V
i
I
550 mA
i
P
5.5 W
i
C
5 nF
i
L
10 μH
i
• Configuration is retained in the HistoROM (EEPROM).
• Error messages (incl. value of operated hours counter) are stored.
No special measures for potential equalization are required.
If the device is designed for hazardous areas, observe the information in the documentation
"Safety Instructions" (XA).
• Without integrated overvoltage protection
Plug-in spring terminals for wire cross-sections 0.5 to 2.5 mm
• With integrated overvoltage protection
Screw terminals for wire cross-sections 0.2 to 2.5 mm
Connection of power supply and signal line
To be selected in feature 050 "Electrical connection"
• Gland M20; Material dependent on the approval:
– For Non-Ex, ATEX, IECEx, NEPSI Ex ia/ic:
Plastics M20x1.5 for cable ⌀5 to 10 mm (0.2 to 0.39 in)
– For Dust-Ex, FM IS, CSA IS, CSA GP, Ex nA:
– For Ex d:
No gland available
• Thread
– ½" NPT
– G ½"
– M20 × 1.5
• Plug M12 / Plug 7/8"
Only available for Non-Ex, Ex ic, Ex ia
Connection of remote display FHX50
Feature 030 "Display, Operation"
L: "Prepared for display FHX50 + M12 connection"
M: "Prepared for display FHX50 + M16 gland, custom connection"
N: "Prepared for display FHX50 + NPT1/2 thread, custom connection"
• Devices without integrated overvoltage protection
Pluggable spring-force terminals for wire cross-sections 0.5 to 2.5 mm
• Devices with integrated overvoltage protection
Screw terminals for wire cross-sections 0.2 to 2.5 mm
• For ambient temperature T
2
2
≥60 °C (140 °F): use cable for temperature T
U
2
(20 to 14 AWG)
(24 to 14 AWG)
Cable entry for FHX50 connection
M12 socket
M12 cable gland
NPT1/2 thread
2
(20 to 14 AWG)
(24 to 14 AWG)
+20 K.
U
35
Need help?
Do you have a question about the Levelflex FMP55 and is the answer not in the manual?