User Calibration And Preparations; Overview; Calibration Modes - Deepace KC901M User Manual

Table of Contents

Advertisement

and Preparation before Measurement

3.1 Overview

KC901M has been calibrated on its ports at factory already. Due to various
reasons, there may have deviations where the displayed curves may appear drifting
and not necessary reflect the exact situations in some cases. Yet in common
engineering applications, the deviations would not tarnish judgments thus there is no
need to conduct another calibration.
In terms of measurement, cables and adapters are commonly used but these
accessories will induce loss and phase shift. Furthermore, its performance is subject
to changes such as temperature and time, accompanied with drift. Under the
circumstances which interferences could seriously impact the results of the measure,
the instrument would need an immediate calibration.
Calibration data is only valid if measuring DUT under the same conditions (cables,
adaptors, temperature, analyzing bandwidth and sweeping speed etc.) as performing
the calibration; or else new errors will occur when any condition changes. In a word,
calibration conditions should be as close as it can be the actual situation of the test.
Try to reduce the frequency of changing different adaptors during testing,
particularly it's not recommended to put on an extra adaptor for whatever reason.

3.2 Calibration Modes

There are three kinds of calibrations in KC901M, factory calibration, system
calibration and user calibration.
Factory calibration: Instrument has a high-density full-band calibration, using
scaling algorithm to conduct more precise interpolation. The factory calibration is
saved into the instrument permanently; users cannot reset it or use it directly. But it
can override the system calibration by restoring factory settings.
System calibration: after the instrument comes out from the factory, the data of
system calibration is the same as factory calibration. System calibration can be used
directly and recalibrated by the users. It avoids further calibration since the System
Calibration in that in whichever frequency the device will execute sbasic error
correction. If it runs the "factory reset", the system calibration which is set by user will
be abandoned and the system calibration data will be the same as the factory
calibration.
User calibration: it can be set by user anytime. This calibration will do the
correction to every data point according to the actual setting. There will not be
Chapter 3 User Calibration
- 17 -