User Calibration And Preparations; Overview; Calibration Modes - Deepace KC901V Manual

Table of Contents

Advertisement

Chapter 3 User Calibration
and Preparation before Measurement

3.1 Overview

KC901V has been calibrated on its ports at factory so that the field calibration
can be skipped. Due to various reasons, the factory calibration can't correct error in
field-use perfectly, so the curves will have some imperfect drifts. But in normal engi-
neering applications, as long as the testing result can help with analyzing and judging,
there is not necessary to spend time on user calibration.
During measurement, cables and adapters are normally used. They will intro-
duce loss and phase shift. With temperature drift and aging, instrument's perfor-
mance will degrade. If this has big influence on your measurement and makes test
result no longer usable, a user calibration should be performed.
Calibration data is only valid if measuring DUT with the same conditions (cables,
adaptors, temperature, analyzing bandwidth and sweeping speed etc.) as performing
the calibration. If these condition has changed, there will be new errors. Calibration
conditions should be as close as it can be the actual situation of the test.
Try to reduce the times of changing different adaptors during testing. It's not
recommended to put an extra adaptor on just for protection the ports.

3.2 Calibration Modes

There are three kinds of calibrations in KC901V, factory calibration, system cali-
bration and user calibration.
Factory calibration: Instrument has a high-density full-band calibration. using
scaling algorithm to do the interpolation reaching higher density. The factory calibra-
tion is saved into the instrument permanently; users can't reset it or use it directly.
But it can override the system calibration by restoring factory settings.
System calibration: after the instrument comes out from the factory, the data of
system calibration is the same as factory calibration. System calibration can be used
directly. It also can be done by users. After doing the system calibration, every fre-
quency of the instrument is calibrated, the basic error corrections can be maintained.
So it's not necessary to do this calibration again in most situations. When restoring
the factory settings, the system calibration which is set by user will be abandoned and
the system calibration data will be the same as the factory calibration.
User calibration: it can be set by user anytime. This calibration will do the cor-
rection to every data point according to the actual setting. There won't be interpola-
tion errors. Therefore, its accuracy is higher than factory calibration as well as system
calibration. User calibration is only applicable for the set frequency. If the setting pa-
rameters are changed, the calibration data will be useless but still saved temporarily.
If we go back to the last setting parameters, the calibration data is still effective.
- 16 -