The file will be automatically named by system. If you want rename file use the
combined key SHIFT+3.
While connecting the instrument to PC via USB, the instrument is detected as a
USB flash drive. In this case, the SD card is already taken over by the PC, therefore
when connecting to PC, data can't be saved.
In S11, if you need to save the files of S1P format or all original testing points,
you need to save it in the Smith chart mode.
2.11
More Information about the System Settings
Normally the FUNC button pops you into the operating surface, where you can
further access the Advanced System Setting Surface (ASSS) with the combined key
SHIFT +7 . In the ASSS the parameters must be examined to be correct to achieve
measurement validity. You can reset all parameters to the defaults via the Factory
Reset function. But be
Sweep point 450 is recommended to boost up the sweep speed
(1) The sweep speed shall better be MEDIUM, since HIGH level is only applicable
for qualitative analysis because of the louder noises and data deviation under
this mode; and LOW level is ideal for enhancing accuracy though, it requires
minor calibration manually.
(2) Select "INT" for both FREQ REF and TRIGER
(3) Select "INT" for S11 BRIDGE for a built-in bridge
- 16 -