Output B, terminals 66/67:

The output B setting must be selected according to application specific requirements (default
setting = Alarm).
Preset to alarm indication:
Example: If track 1 is not measuring, a
display menu 4, and relay output terminals switch to Off.
Call up indication:
When output B is configured as Call-up, the output is activated by an alarm condition and
remains on until it is manually reset via communication interface and the PDM program (or via
manually power down of the device).
A new alarm will not activate a Call-up function if the Call-up function is still active from a
previous alarm.
6.4.8

Checking the operation readiness

All parameters are now set and defined according to the application.
1. Select: [Device → Download to device...] to download the parameters to the device.
Note
Before downloading the parameters, check that all listed data are loaded or changed and
in accordance with the application requirements.
2. Click [Start] to download all changes from the table to the device.
FUS380/FUE380
Operating Instructions, 09/2016, A5E00730100-AB
6.4 Commissioning via PDM
appears on display. Failure code F1 appears in
Commissioning
57