Sterilization Recommendations

Steam sterilization has been found both safe and
effective for the sterilization of Brasseler USA®
BSPMAX™ II Large Bone Power System handpieces,
attachments and battery packs.
The instruments are capable of withstanding the
recommended exposure times and temperatures of
steam sterilization.
Warnings:
• The use of disinfecting solutions for an exterior
instrument wipe will not sterilize equipment and is
not recommended.
Cautions:
• DO NOT sterilize handpieces with Ethylene
Oxide (ETO).
• DO NOT sterilize handpieces in a STERIS
STERRAD
®
System or comparable sterilization
methods.
• DO NOT sterilize handpieces in cold sterilization
like CIDEX
®
.
• DO NOT "peel pack" handpieces or attachments
for sterilization. Sterilization in a sealed pouch traps
moisture which can cause damage.
• DO NOT sterilize battery pack while connected to
the handpiece.
• Do not Flash Sterilize. It can decrease the life of
the device and may cause premature failure of the
device. Flash sterilization will void the warranty.
16
Notes:
• These processes have been validated as being
capable of cleaning and sterilizing the BSPMAX™ II
Sagittal/Oscillator Saw.
• The sterilizer manufacturer's written instruction for
cycle parameters, load configuration and AAMI
guidelines for steam sterilization should be
followed.
®