Care & Maintenance
Brasseler USA® Surgical Power & Accessories
recommends that all BSPMAX™ II components
(handpieces,
attachments
excluding battery packs) be returned to Brasseler
U.S.A. Medical, LLC, Service Department for routine
preventative maintenance every twelve (12) months.
Follow a regular care regimen that includes routine
cleaning and a thorough inspection for damage.
Routine preventive maintenance performed every
twelve (12) months by the Brasseler U.S.A. Medical,
LLC, Service Department can increase the reliability and
extend the life of your BSPMAX™ II Large Bone Power
System.
Checklist:
• Inspect all BSPMAX™ II components (handpieces
and accessories) to verify all components are
present.
• Ensure all handpieces and accessories have been
properly cleaned as outlined in the Cleaning
Recommendations section.
14
and
accessories
• Inspect all devices for proper set up and operation.
• Ensure the battery pack installs properly onto the
handpiece.
• Ensure the saw blade installs properly.
• Ensure there are no loose or missing components.
• Ensure there are no cracks in the battery pack case.
• Test* the handpiece and battery pack to ensure
that they are working properly.
• Check all moving parts for free movement.
• Check for unusual sounds or vibrations.
• Check for proper operating speed.
• If the recommended solutions provided in the
Troubleshooting section do not solve problem(s),
send the device(s) to Brasseler U.S.A. Medical, LLC,
Service Department as outlined in the Repair Service
section.
• Remove battery pack from handpiece when test is
completed.
* This maintenance test may be completed under
non-sterile conditions with a battery pack that has
not been sterilized.