Cleaning Recommendations

Warnings:
• Clean and sterilize handpieces, battery packs and
accessories before every use.
• Prior to cleaning and sterilization, remove saw
blade and battery pack from the handpiece.
• DO NOT use solvents, lubricants, or other
chemicals, unless otherwise specified.
Cautions:
• Follow universal precautions and protective apparel
when handling and cleaning contaminated
instruments.
• Dispose of all saw blades properly after each use.
• DO NOT lubricate handpieces.
• DO NOT immerse handpieces and battery packs
in liquid. Contaminants will enter the equipment
and damage the device.
• DO NOT clean handpieces or battery packs in an
automated washer or ultrasonic cleaner.
• DO NOT clean handpieces with bleach, chlorine
based detergents, liquid or chemical disinfectants, or
any products containing sodium hydroxide (i.e.
INSTRU-KLENZ
®
or Buell
degrade the anodized aluminum coating.
®
Cleaner). They will
Cleaning Procedures:
1. Remove battery pack and saw blade from the
handpiece.
2. Scrub debris from the handpiece using a brush
with stiff, non-metallic bristles and mild pH
balanced enzymatic cleaner. Manipulate all
moving parts of the handpiece to ensure all
debris is removed.
3. Rinse all external surfaces of the handpiece
under running tap water. Hold the handpiece
upright to prevent water from running into the
battery receptacle.
4. If water leaks into the handpiece, hold the
handpiece upright to allow drainage from the
battery receptacle area.
5. Visually inspect the handpiece for any remaining
debris. If any debris is present, repeat the
cleaning and rinsing procedure.
6. Gently shake the handpiece free of water.
7.
Dry the handpiece with a clean lint-free soft cloth.
8. Inspect
handpiece
malfunctioning. Return damaged components to
Brasseler U.S.A. Medical, LLC, Service Department.
9. Place handpieces into designated locations in the
tray and place tray into sterilization case. Fasten
sterilization case lid.
10. S t e r i l i z e a s d i r e c t e d . S e e S t e r i l i z a t i o n
Recommendations section.
for